AMMRL:

From: Kirk Manfredi <kirk.manfredi_at_uni.edu>
Date: Wed, 10 Oct 2018 18:27:43 -0500

Our campus had a university wide power spike last night (actually two of them). The result is that it took down our LC-MS and our 400 Inova (23 shim set 2004 console). After rebooting the system I could never get shims or Z0 to load. With the help of Dave vander velde, I am pretty sure that the problem is the Shim CPU board (based on the symptoms I described to Dave and the two lit LED’s). I can get the board from Agilent (part number 0190299603) but there is no return if it is not the problem. Would anyone have an idle 400 inova with that board I could use to confirm our diagnosis before I purchase it?

Thanks in advance,
Kirk Manfredi
University of Northern Iowa
Received on Wed Oct 10 2018 - 13:27:55 MST

This archive was generated by hypermail 2.4.0 : Tue Jun 20 2023 - 16:35:20 MST