RE: AMMRL: Helium Gas Specification for Refilling Purpose

From: Manoj Naik <manonaik_at_gmail.com>
Date: Sun, 12 Oct 2014 15:05:24 +0530

Dear John,

           Thanks for the information.

It is good device, but it still require a high pressure Helium gas cylinder.
Pressuring Liquid Helium container by a bag using its own low pressure gas
with addition of external heat
doesn't require
​
any additional high pressure cylinder and eventually eliminating issue of
adding the impurities to Liquid Helium container.
​ That's I feel is big advantage.

​Cheers,

Manoj​
 ​

On 11-Oct-2014 12:34 AM, "Jon Webb" <jon_at_mrr.com> wrote:

> Hi Manoj,
>
>
>
> Hi Jerry,
>
>
>
> I hope all is well these days. MR Resources does have a kit on our web
> site for pressurizing LHe dewars via a compressed gas helium tank. Comes
> with all the necessary fittings, tube, regulator and gauge. I strongly
> recommend this over the bag method. If you would like to pass this along to
> some of your colleagues, they may appreciate it.
>
>
>
> Thanks much.
>
>
>
>
>
> Best regards,
>
>
>
> Jon
>
>
>
> www.mrr.com
>
>
>
> MR Resources, Inc.
>
> The Leader in NMR Systems, Services and Accessories
>
> 160 Authority Drive
>
> Fitchburg, MA 01420
>
> 978-345-9011
>
> 1-888-4-NMR-HELP
>
>
> ------------------------------
>
> *From:* Manoj Naik [mailto:manonaik_at_gmail.com]
> *Sent:* Wednesday, October 08, 2014 2:59 AM
> *To:* Matthias Findeisen; AMMRL_at_ammrl.org
> *Subject:* Re: AMMRL: Helium Gas Specification for Refilling Purpose
>
>
>
> Dear Mathias,
>
>
>
> Please check the link below for images & video clip of a device to
> pressurize the Helium container. It require to paddle up (please see the
> video clip) several time to build the pressure in a Helium container, which
> may be not that convenient as that of Helium gas cylinder, but it make
> Helium gas cylinder redundent. Free of impurity concern & saving a
> inventory of Helium gas cynlider.
>
>
>
>
> https://plus.google.com/photos/107209094718337103609/albums/6067329649340496785?cfem=1&authkey=CMq-yMyEiomCngE
>
>
>
> Hope that this could be a usefull device in NMR lab.
>
>
>
> Cheers,
>
> Manoj
>
>
>
>
>
> On Fri, Oct 3, 2014 at 12:10 AM, Manoj Naik <manonaik_at_gmail.com> wrote:
>
> Dear Mathias,
>
>
>
> Helium gas 99.996 is perfectly fine for pressuring Liquid He
> container.
>
> We had used same grade for more than 10 years.
>
>
>
> But since last year we have devised another mechanism to pressurize the
> container.
>
> We are using Hot water bag(available with chemist shop) with adaptor and
> Silicon tube to connect it
>
> at the Vent port (where you connect Helium Gas) of Helium container. This
> bag is sandwich in
>
> wooden folded planks & we paddle it to increase the container pressure.
> So no more need of Helium Gas cylinder.
>
> Unfortunately, right now I am not able to send you the images & video
> clip of our device as I am away, but
>
> surely I will share that device with AMMRL community when I will be back
> to NMR Facility.
>
>
>
> Cheers,
>
> Manoj
>
>
>
> On Thu, Oct 2, 2014 at 2:59 PM, Matthias Findeisen <
> matthias.findeisen_at_uni-leipzig.de> wrote:
>
>
> Dear all,
>
> which Helium-Gas Cylinder do you use for refilling the magnet (i.e. to
> give slight pressure onto the refill dewar)?
>
> Is the specification of 99.996 % sufficient for this?
>
> Which experiences do you have?
>
> Thanks!
>
> Yours
> Matthias Findeisen
>
> Univ- Leipzig/ Germany
>
>
>
>
>
> --
>
> *Manoj Vishnu Naik, *
>
> NMR Facility,
>
> *Tata Institute of Fundamental Research, *
>
> Homi Bhabha Road, Colaba,
> Mumbai, INDIA-400005
> Tel: +91 22 22782525
> Fax:+91 22 22804682
>
>
>
>
>
> --
>
> *Manoj Vishnu Naik, *
>
> NMR Facility,
>
> *Tata Institute of Fundamental Research, *
>
> Homi Bhabha Road, Colaba,
> Mumbai, INDIA-400005
> Tel: +91 22 22782525
> Fax:+91 22 22804682
>
> No virus found in this message.
> Checked by AVG - www.avg.com
> Version: 2015.0.5315 / Virus Database: 4181/8353 - Release Date: 10/09/14
>
Received on Sat Oct 11 2014 - 23:35:26 MST

This archive was generated by hypermail 2.4.0 : Sun Jun 18 2023 - 18:51:47 MST