Everyone:
I have suddenly lost my shims on our VNMRS-500 system. I don't think
that it is the cryoshims because the lineshape is ok with all of the
rt shims set to zero. Varian agrees that we have probably lost a rt
shim. I need some advice in how to trouble shoot the shim power
supply. Varian says that I cannot trouble shoot the shims without
buying a resistor board from them for testing for the current. This
board is $1000 and I would prefer not to buy it just yet. I would
appreciate any help in this matter.
Thanks.
Susan
Susan L. Nimmo, Ph.D.
NMR Facility Manager
Department of Chemistry and Biochemistry
University of Oklahoma
101 Stephenson Parkway
Norman, Oklahoma 73019
office: 405-325-0677
cell: 405-627-7044
lab: 405-325-2843
Received on Wed Oct 27 2010 - 06:00:40 MST