Hi everyone,
Given the state of helium supply and prices within the next 10 years I'm considering some different operating procedures to try to capture more of my venting helium. The University of Alberta has a campus wide helium recovery system which runs through 5 different building and 3 different departments. In the past helium applications fed into the recovery lines and go back to a liquefaction plant in the Physics department.
That system was over 30 years old and hasn't been in operation for 8 years, but a new system has been purchased and helium recovery and liquefaction will resume early next year.
My question is does anyone have any experience in trying to capture venting helium gas during a helium fill on an NMR? I know nominal operating pressures are just above atmospheric pressure and suppliers don't recommend significant pressurizing of the helium can. Originally the helium recovery line was a mere 2" of water column (0.07psi), so the venting pressure of 1 to 2 psi would allow the gas to flow nicely. Low pressure helium recovery would have to be substantial, or compression and storage of recovered low pressure gas must be fast.
I would like to hear your thoughts. I've also heard Bruker will void your warranty if you hook a Bruker NMR up to helium recovery. The suppliers are going to have to start supplying solutions to helium recovery in the not to distant future.
I will absolutely post a summary,
Thank you.
Deryck
------------------------------------------------------------------------------------
Deryck Webb - Business and Technical Manager
NANUC - Canada's National High Field NMR Centre
Room 103 Email: deryck_at_nanuc.ca
NANUC Bldg. Phone: (780) 492-8530
University of Alberta Fax: (780) 492-9174
Edmonton, Alberta Cell: (780) 707-0857
Canada T6G 2E1 AIM: derycknmr
http://www.nanuc.ca
Received on Fri Oct 15 2010 - 13:13:39 MST