O2 level sensor

From: LAWRENCE BYRNES <numare_at_eclipse.net>
Date: Thu, 12 Oct 100 16:13:42 PDT

Hello again Spinlanders,

FedEx delivered a Crowcon Gasman II oxygen level sensor today in time for a
liquid nitrogen fill. During the course of the twenty four minute fill, the
oxygen level at my desk fell from 20.8% to 20.0% v/v. We have an 18,000 BTU
wall mounted air conditioner that runs maxed out 24/7 all year long. The lab
is ca. 16' x 18' x 10'. I suspect that laboratories with less ventilation
per cubic foot will observe more profound drops in oxygen levels during
routine fills.

Just thought that some of you folks might be interested.

Kind regards,

Lawrence
--------------------------------------------------------
10/12/00 16:13:42
Lawrence Byrnes,
NMR Laboratory Manager,
Numare Spectralab Incorporated
Research Director,
The Schoenenberg Institute of Molecular Spectroscopy
10 Summit Avenue, Building #6, P.O. Box 96,
Berkeley Heights, NJ, 07922-0096,
telephone 908-665-0066
facsimile 908-464-4192
email numare_at_eclipse.net
www http://www.eclipse.net/~numare/
--------------------------------------------------------
Received on Thu Oct 12 2000 - 20:14:07 MST

This archive was generated by hypermail 2.4.0 : Sat Jun 03 2023 - 17:40:01 MST