RE: AMMRL:Summary of Varian VNMRJ on PC with Linux

Sardashti, Mazi (Mazi.Sardashti@conocophillips.com)
Mon, 26 Nov 2007 14:16:22 -0600


This is a multi-part message in MIME format.

------_=_NextPart_001_01C83069.3683DF40
Content-Type: text/plain;
	charset="us-ascii"
Content-Transfer-Encoding: quoted-printable

This is a favor I need to ask to help another AMMRL member.
Unfortunately, I have lost the contact information from the original
responses to my question about the Sun upgrade to a PC with Linux on my
INOVA. This is because our mail system automatically deletes anything
older than 30 days old. So, could the person who sent me the response
no. III below, send me a note so that I can forward your contact info to
the person who has a specific question for you.
=20
Thanks in advance,
Regards,
Mazi.=20
 =20
Maziar Sardashti=20
ConocoPhillips Co.=20
130 PL, BTC=20
Bartlesville, OK 74004=20
Phone: 918-661-4129=20
FAX: 918-662-2888=20
mazi.sardashti@conocophillips.com
 =20
 =20


________________________________

From: Sardashti, Mazi=20
Sent: Tuesday, October 30, 2007 8:40 AM
To: AMMRL
Subject: AMMRL:Summary of Varian VNMRJ on PC with Linux



Dear AMMRL colleagues,=20
Thank you very much to all who responded to my question on using PC with
Linux and VNMRJ. I have attached a complete list of responses below. In
summary, here is what I'll have to worry about:

1. Make sure the Linux installed is the correct version, and the Dell
model/setup is the same as what is recommended by Varian.

2. Find out the data acquisition processor type in the RF cabinet: power
PC or Motorola.=20
3. The ACC command for running the Chemagnetics solids spinner
controller will NOT work on the PC. I will have to keep the Sun computer
to run solids.

As you may expect, I am not very happy about #3. The main reason that I
am doing the upgrade is to get rid of an old computer/operating system
that is out dated and not supported anymore. If I have to keep the Sun,
even for just running the solids, then it will defeat the purpose.

So, I am still thinking about whether I should go through with the
upgrade or not. If I decide to do it, I'll post any additional issues
not covered by others after the installation.

Thanks again for all your help. As I always say, AMMRL is an invaluable
resource and a great group to be a part of.=20

Best wishes,=20

Mazi.=20
 =20
Maziar Sardashti=20
ConocoPhillips Co.=20
130 PL, BTC=20
Bartlesville, OK 74004=20
Phone: 918-661-4129=20
FAX: 918-662-2888=20
mazi.sardashti@conocophillips.com
 =20
 =20
I. I went through a similar upgrade a few years back.  Now I'm working=20
exclusively with Bruker systems, however. =20

You mentioned that the Linux is coming pre-installed.  Just a tip - if I

were you I'd reinstall Linux from scratch. It's a straight-forward=20
operation.  Make sure to install ALL features of the operating system
before=20
you install VNMRJ.  With an INOVA 'hooking' up to the console should be
as=20
simple as plugging in - assuming you have an Ethernet card for both the=20
world and the spectrometer.=20
___________________________________=20

 II.  You should be aware that the "Acc" software that runs the
Chemagnetics=20
spin controller WILL NOT (according to the latest information that I
have=20
from Varian) run under Linux. So, even if you upgrade the host computer
to=20
Linux, you will have to keep a Sun around to run the spin controller.=20

_____________________________________=20
 III. I can't help you with the robot. As for the rest, my suggestions
are:=20

1. Make sure the 390N has the configuration Varian suggests. Mostly this

means get the pair of Broadcom network cards and the "correct" video
card.=20
The hard drives, CPU speed, and RAM shouldn't matter.=20

2. Can you receive Word documents as attachments? I can send you a short

document I wrote up with my experiences getting an Inova to work this
way.=20
It's not comprehensive, but might save you some time/headaches. There
are a=20
few things with getting the network to behave, and if your Inova is old=20
enough to have a M68k CPU as oppose to PowerPC, there are a few extra
things=20
you have to do (I'm pretty sure the last part can be found on Varian's
user=20
pages as well).=20

3. I would suggest seeing if your Varian rep will give you a copy of=20
Varian's modified RHEL4 install DVD. There are a few extra packages that
you=20
need that aren't there by default which may or may not be on the VJ
Linux=20
install CD. If not, make sure you do a full Linux install (ALL packages)
and=20
I have the additional rpms. If you can get the Varian disk, I have some=20
instructions to modify the kickstart file (for your own root password,=20
network info, etc.) that I can tar up and send you.=20

The configuration you're planning can work. Good luck!=20
_______________________________________=20
 IV. I upgraded an Inova console to VJ 2.1B.  The only stumbling block
was=20
the initial communication and download of the diskless kernel to the=20
CPU.  The fix and work around was outlined in a bug report listed in the

Read me file for 2.1B. =20

I do not do solid experiments with this console, but we do have an=20
SMS-50 that works fine.=20
___________________________________________=20
 V.  I would double-check with Varian,but the latest info that I have=20
is that acc wouldn't run on Linux.  Now,  if they get enough=20
pressure from customers, maybe they will do a Linux port of acc. =20

I think that with the new "Direct Drive" consoles that the spin
controller is built=20
into the console. and communicates with the Linux computer through the=20
console.=20
___________________________________________=20
 VI.  Mazi - Do you have a powerPC or a Motorola controller in your
inova's=20
RF cabinet?   We bought the PC and vnmrj before anyone realized our=20
motorola controller didn't work with vnmrj.  It took a lot of work by=20
a 3'rd party to make it go.    Check with Varian about compatibility,=20
and whether any PowerPC controllers have become available if you need
one.=20

Check whether the Dell is preinstalled with 64-bit Linux.  Mine came=20
with 32-bit Linux and the difference wasn't clear until=20
troubleshooting why my  VJ installation wasn't working.   Installing=20
64-bit Linux was straightforward.=20

______________________________________________=20





------_=_NextPart_001_01C83069.3683DF40
Content-Type: text/html;
	charset="us-ascii"
Content-Transfer-Encoding: quoted-printable


AMMRL:Summary of Varian VNMRJ on PC with =
Linux



This=20 is a favor I need to ask to help another AMMRL member. Unfortunately, I = have=20 lost the contact information from the original responses to my question = about=20 the Sun upgrade to a PC with Linux on my INOVA. This is because our mail = system=20 automatically deletes anything older than 30 days old. So, could the = person who=20 sent me the response no. III below, send me a note so that I can = forward=20 your contact info to the person who has a specific question for=20 you.
 
Thanks in advance,
Regards,

Mazi. =
 
Maziar Sardashti
ConocoPhillips Co.
130 PL, BTC
Bartlesville, OK 74004
Phone: = 918-661-4129
FAX: 918-662-2888 =
mazi.sardashti@conocophillips.com
 =20



From: Sardashti, Mazi =
Sent: Tuesday,=20 October 30, 2007 8:40 AM
To: AMMRL
Subject: = AMMRL:Summary of=20 Varian VNMRJ on PC with Linux

Dear AMMRL colleagues, =
Thank you very much to all who responded to my = question=20 on using PC with Linux and VNMRJ. I have attached a complete list of = responses=20 below. In summary, here is what I'll have to worry about:

1. Make sure the Linux installed is = the correct=20 version, and the Dell model/setup is the same as what is recommended by=20 Varian.

2. Find out the data acquisition = processor type=20 in the RF cabinet: power PC or Motorola.
3. The ACC command for running the Chemagnetics = solids=20 spinner controller will NOT work on the PC. I will have to keep the Sun = computer=20 to run solids.

As you may expect, I am not very happy = about #3.=20 The main reason that I am doing the upgrade is to get rid of an old=20 computer/operating system that is out dated and not supported anymore. = If I have=20 to keep the Sun, even for just running the solids, then it will defeat = the=20 purpose.

So, I am still thinking about whether = I should=20 go through with the upgrade or not. If I decide to do it, I'll post any=20 additional issues not covered by others after the = installation.

Thanks again for all your help. As I = always say,=20 AMMRL is an invaluable resource and a great group to be a part = of.

Best wishes,

Mazi.
 =20
Maziar Sardashti
ConocoPhillips Co.
130 = PL, BTC=20
Bartlesville, OK 74004
Phone: 918-661-4129
FAX: = 918-662-2888
mazi.sardashti@conocophillips.com 
  
I. I = went through a=20 similar upgrade a few years back.  Now I'm working
exclusively with Bruker systems, = however. =20

You mentioned that the Linux is = coming=20 pre-installed.  Just a tip - if I
were you I'd reinstall Linux from scratch. It's a = straight-forward=20
operation.  Make sure to = install ALL=20 features of the operating system before
you install VNMRJ.  With an INOVA 'hooking' up to the = console should=20 be as
simple as plugging = in -=20 assuming you have an Ethernet card for both the
world and the spectrometer.=20
___________________________________

 II. =  You=20 should be aware that the "Acc" software that runs the Chemagnetics=20
spin controller WILL NOT = (according=20 to the latest information that I have
from Varian) run under Linux. So, even if you upgrade the host = computer=20 to
Linux, you will have = to keep a Sun=20 around to run the spin controller.

_____________________________________ 
 III. I can't = help you=20 with the robot. As for the rest, my suggestions are:

1. Make sure the 390N has the = configuration=20 Varian suggests. Mostly this
means=20 get the pair of Broadcom network cards and the "correct" video = card.=20
The hard drives, CPU speed, and = RAM=20 shouldn't matter.

2. Can you receive Word documents = as=20 attachments? I can send you a short
document I wrote up with my experiences getting an Inova to = work this=20 way.
It's not = comprehensive, but=20 might save you some time/headaches. There are a
few things with getting the network to = behave, and if=20 your Inova is old
enough = to have a=20 M68k CPU as oppose to PowerPC, there are a few extra things =
you have to do (I'm pretty sure the last = part can be=20 found on Varian's user
pages as=20 well).

3. I would suggest seeing if your = Varian rep=20 will give you a copy of
Varian's=20 modified RHEL4 install DVD. There are a few extra packages that = you=20
need that aren't there by = default which may=20 or may not be on the VJ Linux
install=20 CD. If not, make sure you do a full Linux install (ALL packages) = and=20
I have the additional rpms. If = you can get=20 the Varian disk, I have some
instructions to modify the kickstart file (for your own root=20 password,
network info, = etc.) that I=20 can tar up and send you.

The configuration you're planning = can work.=20 Good luck! =
_______________________________________ 
 IV. I upgraded an Inova = console to VJ=20 2.1B.  The only stumbling block was
the initial communication and download of the diskless kernel = to=20 the
CPU.  The fix = and work=20 around was outlined in a bug report listed in the
Read me file for 2.1B. 

I do not do solid experiments = with this=20 console, but we do have an
SMS-50=20 that works fine.=20
___________________________________________ 
 V.=20  I would double-check with Varian,but the = latest info=20 that I have
is that acc = wouldn't run=20 on Linux.  Now,  if they get enough
pressure from customers, maybe they will = do a Linux=20 port of acc. 

I think that with the new "Direct = Drive"=20 consoles that the spin controller is built
into the console. and communicates with the Linux computer = through=20 the
console.=20
___________________________________________ 
 VI.=20  Mazi - Do you have a powerPC or a Motorola = controller=20 in your inova's
RF=20 cabinet?   We bought the PC and vnmrj before anyone realized = our=20
motorola controller = didn't work with=20 vnmrj.  It took a lot of work by
a 3'rd party to make it go.    Check with Varian = about=20 compatibility,
and = whether any=20 PowerPC controllers have become available if you need one.

Check whether the Dell is = preinstalled with=20 64-bit Linux.  Mine came
with=20 32-bit Linux and the difference wasn't clear until
troubleshooting why my  VJ = installation wasn't=20 working.   Installing
64-bit Linux was straightforward.

______________________________________________=20