---- Richard Shoemaker, Ph.D. Phone--(402) 472-6255 Instrumentation Director, Chemistry FAX---- -6964 Research Associate Professor, Chemistry University of Nebraska-Lincoln URL: http://www.chem.unl.edu/nmrlab.html -----Original Message----- From: John Snyder [mailto:john-snyder_at_uiowa.edu] Sent: Wednesday, November 01, 2000 7:37 AM To: ammrl_at_wwitch.unl.edu Subject: Re: 31P-1H: Thanks for info Knowing the Gradient ratios to use for 31P will be useful. Is there a comprehensive source for Gradient Ratios for some of the other nuclei other than 13C, 15N, and 31P? At 12:12 PM 10/31/00 -0800, you wrote: >Thanks to the many people who offered helpful suggestions for performing >1H-31P long range correlation experiemnts. > >The most important advice was to set the z-gradient ratio to 46.77:60:30, >which produced a lovely hmbc spectrum using a delay of 70 ms. > >Thanks again, > >Phil. > >-- >Phil Dennison >NMR Facility Manager >UC Irvine > > > ------------------------------------------------------------- John Snyder email: john-snyder_at_uiowa.edu NMR Facility ph: 319-335-1332 Chemistry Dept. Dept.fax: 319-335-1270 University of Iowa Iowa City, IA 52242 --------------------------------------------------------------