spurious signals

Stefan Berger (stberger@rz.uni-leipzig.de)
Wed, 1 Dec 1999 16:01:14 +0000

I would like to add my experience to the discussion. Several yars ago
we had the same problem on an AM-400. The source of the signal at
about 170 ppm was traced to a nearby (3 meter) X-32 data station.
Replacement of some oscillator on one of its boards was the remedy to
the problem.

stb
-------------------------------------------------
Prof. Dr. Stefan Berger
Institut fuer Analytische Chemie
Universitaet Leipzig
Linn=E9str.3
04103 Leipzig

Tel: +49 341 9736101
Fax: +49 341 9711833 or 9736115
e-mail: stberger@rz.uni-leipzig.de
home page: http://www.uni-leipzig.de/~nmr/STB/stb.html