Subject: Z0 coil field strength survey
We have discovered (on a new Varian Inova AS600 instrument) that the
field strength
of our Z0 shim coil is about 22000 Hz total (+/- 11000 Hz). This means
that for a
relatively stable magnet, drifting at 2 hz/hour, the Z0 coil will not be
able to maintain
the field, and lock frequency, beyond around 11000 hours {(22000
Hz)/(2Hz/hr)} which is
a little over a year. So, the spectrometer frequency has to be changed
roughly every year.
In my experience, this is rather unusual. Varian insists that this is
perfectly normal, and
has always been so. I am trying to gather information about the
'typical' field strength
of a high field spectrometer Z0 coil. I would be grateful to anyone who
would be willing
to supply this information, in any form (i.e. a number in Hz, or how
ofter do you need
to change the spectrometer frequency on your system), along with the
field drift rate
of the instrument.
I will post a summary of the responses.
Many Thanks,
David Eliezer