Re: [AMMRL] AMMRL: lock problem

From: Martha Morton <mmorton4_at_unl.edu>
Date: Tue, 30 May 2023 16:58:14 +0000

Karel,

Please tell us your ECL level for the 2H-TX board.

Martha

Martha Morton
Director of Research Instrumentation
Co-Chair, Department of Chemistry Safety Committee
University of Nebraska-Lincoln
4024726255<tel:4024726255>

________________________________
> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> on behalf of Karel Klika <karkli_at_utu.fi>
> Sent: Tuesday, May 30, 2023 11:55 AM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] AMMRL: lock problem
.
> Dear all,
>
> Regarding the strange lock problem. many thanks to all those who responded
> with suggestions. Still no closer to a resolution unfortunately.
> I think it is the 2H-TX board, so anyone have such a board for a Avance 600
> they are willing to part with ?

> Also, I should apologize for the "2D" instead of saying "2H" or just "D" for
> example which confused some people (deuterium or 2-dimensional ?), a bad habit
> I picked up from a certain NMR program that I used to use a long time ago.
>
> Many thanks.
>
> Regards,
> Karel
> ________________________________________
> From: Karel Klika
> Sent: 26 May 2023 13:26:12
> To: main_at_ammrl.groups.io
> Subject: AMMRL: lock problem
>
> Dear all,
>
> We have a problem with lock/topshim. We can lock fine on 2D, lock is normal
> and S/N is good. If we run topshim on 1H all is fine. However, topshim on
> 2D fails and we get an error message saying no signal at all. Measuring a 2D
> spectrum only provides a very weak signal. If I try and measure the 2D pulse
> length it looks really weird (pic attached), the signal intensity does not
> seem to change at all, but the phase is totally erratic. Any ideas what the
> problem is and how to fix it ?
>
> Many thanks.
>
> Regards,
> Karel
>

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#157): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/157__;!!PvDODwlR4mBZyAb0!SvEuTF4qJczKCdKVUweEdMrmZTrioh7N=
i0dwzWJL-4fQG_bttofo7r2UxJp-sj3EzDridY386Un9oFzS$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/99149994/=
7559972__;!!PvDODwlR4mBZyAb0!SvEuTF4qJczKCdKVUweEdMrmZTrioh7Ni0dwzWJL-4fQG_=
bttofo7r2UxJp-sj3EzDridY386UdSCRWh$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-


Received on Tue May 30 2023 - 10:09:23 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:41:33 MST