AMMRL: ASX400: ASX32 Computer - no connection - VMC left light diode on steady

From: Michael Fechtelkord <Michael.Fechtelkord_at_ruhr-uni-bochum.de>
Date: Fri, 24 May 2019 15:14:18 +0200

Dear colleagues,


we have have a problem with an ASX32 Computer on a ASX400 console. We
had no 5V sense from the EWS 600 because the cable was tightened on the
screw. In addition we saw some short cut (foam, smell and the 230V fuse
went down). We found the cable and now all voltages are back. But now we
get no connection between SGI O2 and the CPU4 board ECL02 in the X32.
Usually the left yellow diode under the Ethernet connection turns off
after a while but now it is always on. We changed also the CPU4 memory
board against a ECL03 version but the result is the same.


Does anyone have experience with that or a manual or circuit diagrams?
That would help.


Thanks in advance!


Best regards,

Michael Fechtelkord


-- 