I asked about transporting liquid nitrogen to a basement facility
when/if the elevator is not available. As usual, the membership was
pretty creative. Thanks to everyone who responded.

1) Simple and easy: (if outage is planned) Fill the magnets and refill
the dewar. Beg, borrow, or rent additional dewars, fill them, and park
all of them in the basement. Depending on the quality (and quantity) of
the storage dewars, this could keep the magnets full for several weeks.
  (We do this over winter break.)

2) Simple and easy: (if there is basement access). Bring the dewar in
via a loading dock or opening. In our case, there is no loading dock at
all. There is one outside stair that goes down to basement level, and
we have discussed in the past lowering a dewar (with a crane) in an
emergency.

3) Move the dewar downstairs by hand. Riggers and piano movers deal
with large, awkward, heavy things all the time. Professionals could
probably move a full dewar. Even a half-full one (moved 3 times) would
work. (Note: one respondent used this method to move cryogens
UPstairs.) Alternate: put in a ramp on the stairs to move the dewar.

4) Move only the liquid. Leave the dewar at ground level and run a long
hose to the basement to the magnets. If there are windows or other
openings, use them, otherwise run the hose down the stairs. (Or, drill a
hole- we _are_ renovating.) Yes, cool-down will take some nitrogen, but
eventually, liquid will flow. Alternate: Put the storage dewar at the
base of the stairs, and fill it this way, then move to the facility.

5) Pour in nitrogen via a funnel and a small dewar. Mixed response to
this, ranging from don't because spills will freeze the o-rings, to
tedious and prone to splashing, to please video and post.

6) Use small (25 liter) dewars and pressurize them to push the liquid
out. A stopper with 2 holes with 2 tubes. A tube in the bottom connects
to a fill hose. A tube just in the top provides gas access. These can
(almost) be carried up and down the stairs, or could be filled from 10
liter dewars. (I have several drawings and pictures of these setups if
anyone needs them.)

7) Use lawyers. Make the contractors financially responsible for getting
the nitrogen and/or limiting the elevator down time. (Or replacing the
magnets.) Note that it may be possible to get permission to use the
elevator to transport dewars even if it is not fully functional/not yet
inspected/not safe for people.


kas
-- 
Karen Ann Smith           Director, NMR Facility
Group Leader, Analytical Chemistry Services
Department of Chemistry and Chemical Biology
Clark Hall MSC03 2060
1 University of New Mexico           Albuquerque, NM 87131-0001
karenann_at_unm.edu       505.277.4031          http://nmr.unm.edu
Out of this world screensaver: http://setiathome.ssl.berkeley.edu
"Yesterday is history, tomorrow's a mystery, and today is a gift-
that's why we call it the present."  from "The Joy of Stress:
Humor as a Coping Mechanism," by Loretta LaRoche
Received on Mon May 05 2014 - 08:46:42 MST

This archive was generated by hypermail 2.4.0 : Sun Jun 18 2023 - 17:44:46 MST