AMMRL: SCSI cards for Sun Ultra with Varian Unity systems

From: Neil Jacobsen <neil_at_u.arizona.edu>
Date: Mon, 07 May 2007 16:10:31 -0700

Does anyone out there have experience with operating a Unity system
with a Sun Ultra host?

Our Unity-300 has been operating for five years with a Sun Ultra-10,
using the same SCSI cable we used for the previous host (Sparc IPC)
plus a simple adaptor at the Sun end, and a "cheap" SCSI card (Sun
PCI Single Channel Single Ended SE Ultra/Wide SCSI card, part number
X5010A or 375-0097). Recently, in the course of fixing a
communication failure, we tried using the "fancy" SCSI card
recommended by Varian (Sun PCI 10/100 Ethernet and SE Ultra/Wide SCSI
card, part number X1032A or 501-5656). Exactly as Varian predicted
in their original notes on the upgrade, the Sun will not boot with
the "fancy" SCSI card unless the SCSI cable is replaced with a newer
one supplied by Varian (with adaptor at the differential box
end). So we bought the new cable kit from Varian ($280), but the
communication still fails. In the end, we replaced the "cheap" SCSI
card with a new one and, using the old cable, the communication was
restored. We later noticed that a voltage regulator was fried on the
old SCSI card.

What puzzles me is why the "fancy" SCSI card with the new cable kit
from Varian is dead as a doornail. We actually tried two of these
"fancy" cards, so it's unlikely that the card is defective. Does
anyone have any ideas?

Thanks,

         Neil



Neil E. Jacobsen, Ph.D.
NMR Facility Manager
Department of Chemistry
119 Old Chemistry
1306 E. University
University of Arizona
Tucson, AZ 85721
520-621-8146
FAX 520-621-8407
Received on Mon May 07 2007 - 18:12:18 MST