Hi Friends:
We are buying a Bruker AV400 with a 5mm BBI probe. The vibration test was done at the site we are going to put the magnet. The following peaks were observed:
Frequency (Hz) Intensity (mm/s*s)
5 0.35
9 0.25
14 0.25
18 0.50
26 0.35
28 0.45
37 0.65
A Bruker engineer recommended antivibration legs as the intensities are higher than 0.1 and antivibration feet will only reduce vibrations above 8 Hz (the feet will amplify the vibration at 5 Hz).
What kinds of experiments will be affected if we do not put antivibration legs and feet based on your experiences?
Thank you very much for your help.