Due to a short-circuit in the MRI console of a Freiburg hospital (August
2001), heavy smoke penetrated into the adjacent rooms.
Unfortunately the well-trained and competent staff was asked to leave the
area, and so nobody knew that in spite of turning off all power supply the
magnet was still energized.
During a check-up round a fireman with full breathing equipment was drawn
into the bore where he got stuck. Only after quenching the magnet he could
be saved suffering serious injuries.
(see: "Brandschutz" Deutsche Feuerwehr-Zeitung 3/2002 p. 281 ff.)
As a consequence the fire department told us to establish a fixed stoppage
around all >10T-magnets at the campus.