I have in my hand 6.1c --- I was thinking of installing 6.1c and chempack
to make it easier to run hmqc, ghmqc etc. on my U+/HAL, spark10. I was
going to ask the group's experience. Since seeing this email, *now I'm
asking*. :) Cheers, Lew
On Fri, 3 May 2002, Wei Li wrote:
> Please ignore this message if you have not used Varian's chempack 2.1. Recently I installed the new patch (104) for vnmr61C running on my invoa, and installed the new chempack 2.1. I noticed that the appearance of the pulse sequence changed a lot, most noticeably, the delay d2 is replaced by t1. I think those are the updated sequences made from spincad. The problem is whenever I tried to run any 2D experiments (COSY, gHSQC, HMBC, etc) with the new pulse sequence, it always complains that "Invalid Acquistion Instruction" and could not start. The 1D proton is ok. Varian's TAC scientists spent a few days trying to figure out why it did not work on my system, but no luck. I am wondering if anyone there have bumped into the similiar problem and how to solve it. My inova is controlled by an Ultra-10 running Solaris 8 04/01, and we have a PowerPC in the console. I do noticed that when I invoke the macro like gHSQC, I saw a display of " java 1.2.2" in the text window, looks like i!
t uses Java to display the sequence?
>
> Any suggestions will be highly appreciated.
>
> Regards
>
> Wei Li
>
Received on Fri May 03 2002 - 11:40:22 MST