IRIX 6.5 ec0/ec1 problem

From: Wei Wycoff <WycoffW_at_missouri.edu>
Date: Tue, 22 May 2001 14:35:38 -0500

Hi again,

After I have sent the last email, our computer person told me that when he
disabled all the Bruker related process such as gated etc., ec0 connection
to the network would work fine. So the problem now is with ec1 connection
to network (ec0 connection to spect is always fine even when ec1 crashes).

Wei



My last email to all:

Hi all,

We have just installed IRIX 6.5.7 and NMRSuite 2.6/patch level 5 on our
SGI-O2 of ARX250 NMR. We have problem with ethernet connection. The
connection stops when we try to transfer data or read data remotely from
tape drive. Bruker says with IRIX 6.5 and SGI-O2, ec0 should be connected
to CCU and ec1 to internet. We have set it up this way but ec1 network
connection crashed constantly. So we set ec0 back to the network just to
get some old backups read back from tape, that crashed too. We have not
tried to get NMR to work because some old data has yet to be restored from
tape which we can't do due to the crashes. We are thinking about
installing IRIX 6.5.11, but do not know if it will fix the problem. Any
experience and suggestions/solutions are welcome.

Wei

<:> <:> <:> <:> <:> <:> <:> <:> <:> <:> <:> <:> <:> <:> <:> <:> <:> <:>

Wei Wycoff
Department of Chemistry
601 S. College Ave.
University of Missouri-Columbia
Columbia, MO 65211
USA

voice: (573) 882-3291
fax: (573) 882-2754
email: WycoffW_at_missouri.edu
http://chem.chem.missouri.edu/nmr/nmr.html

<:> <:> <:> <:> <:> <:> <:> <:> <:> <:> <:> <:> <:> <:> <:> <:> <:> <:>
Received on Tue May 22 2001 - 18:48:24 MST

This archive was generated by hypermail 2.4.0 : Sun Jun 04 2023 - 15:30:07 MST