We have a problem with 1H decoupling on our AMX400. Usually 25db/100us is
used by zgpg30 (cpd = walt16) to decouple from protons when observing 13C,
but lately it has not been sufficient: for 20%CHCl3 / Acetone-d6 about
80Hz residual coupling is still left. The 1H frequency o2 in this case is
approximately 1.5kHz off-resonance. If this is set on-resonance, splitting
disappears. We have checked both high- and low-power pulses using decp90,
these have not been affected. Have also tried garp and bb modes for cpd,
but splitting is still there.
Any suggestions are most welcome,
Abil