From: celine schneider <cmschnei_at_mun.ca>
Date: Tue, 5 Mar 2019 17:44:59 -0330

Hi,

Our computer crashed and we had to re-install TS1.3 on a “new” box (running XP because our 500 is AV….). Installation went ok, I had a NMR_backup somewhere (used NMR_SAVE), no problem with Flexlm/license, but when doing cf, I’m getting the following:

"connection to spect (aqport0) failed
unable to connect to remote host *Connection refused)
 ('startd' demon active on CCU ?)”

I rebooted the AQS, then the entire console, error is still there (we did that at least 10 times).

CCU is flashing nicely.
I can ping to CCU but I cannot telnet to CCU (it opens and shuts down immediately).
SPEC is in automatic debit (100MBps). If I switched to 10MBps, ping to CCU times out.
All services are started as expected.

Troubleshooting CCU mentioned plugging an Hyperterminal in TTY000, but I cannot find it. We have TTY001-TTY020. no 000.

Any suggestion?

Thank you,

Celine