Re: AMMRL: VT controller has no gas flow

From: Sara D Kunz <kunz_at_brandeis.edu>
Date: Tue, 19 Mar 2013 14:43:41 -0400

Several years ago, after a lot of frustration with Varian VT cables, I
decided to support the VT cable so that it would not hang down putting
stress on the connector and to keep it out harm's way. I wrapped a
bungee cord around one of the magnet legs and arranged the VT cable
under the cord on the inside of the magnet leg and then connected it to
the probe. There was no longer any stress on the connector and the cable
is was out of the way. I no longer get calls that are related to the VT
controller.

Sara Kunz
Brandeis

On 3/18/2013 2:16 PM, Stolowich,Neal J. wrote:
>
> I have noticed this a couple of times on my VNMRS 700 as well. A few
> cycles of resetting the VT and pneumatics, perhaps with rebooting the
> console in between does fix it, but I’m curious as well if there is a
> specific fix, rather than randomly resetting things.
>
> Regards,
>
> Neal
>
> Neal J Stolowich, PhD
>
> NMR Facilities Manager
>
> Department of Chemistry
>
> University of Louisville
>
> *From:*Xuemei Huang [mailto:xhuang_at_ucsd.edu]
> *Sent:* Friday, March 15, 2013 7:26 PM
> *To:* ammrl_at_ammrl.org
> *Subject:* AMMRL: VT controller has no gas flow
>
>
> Hi, Everyone,
>
> We have a 800MHz Varian VNMRS spectrometer equipped with cold probe.
> Recently I am running into a problem where I can not regulate the
> temperature. After I set the temperature, I see the VT starts
> flashing, indicating it's trying to change the temperature. But the
> temperature never changes. I tried to reset the VT controller, the
> pneumatics, as well as resetting the console. None of them seemed to
> solve the problem. The error message I am getting is that : VT
> controller has no gas flow. But I checked that there is sufficient air
> flow. Has anyone seen similar problem on this system? Any suggestions
> would really be appreciated.
>
> Thanks,
>
> Xuemei
>
> ***************************************************
>
> Xuemei Huang
> Department of Chemistry and Biochemistry
> University of California, San Diego
> 9500 Gilman Dr.
> La Jolla, CA 92093
> xhuang_at_ucsd.edu <mailto:xhuang_at_ucsd.edu>
> (858)-822-2307
> ***************************************************
>
>
>
>
Received on Tue Mar 19 2013 - 08:43:44 MST