RE: NMR liquid handler

From: Yan, Jiangli <Jiangli.Yan_at_pfizer.com>
Date: Tue, 28 Mar 2006 11:38:12 -0800

Hi Chris,

We are using a THEONXY robot to transfer solution from a 96-well plate
into NMR tubes routinely. Instead of using a long and sharp handler
which is able to reach to the bottom of NMR tube, we designed a NMR tube
rack. This NMR tube rack has 96 wells and can hold 96 NMR tube steadily.
Horizontally, it has the same size as any other regular 96-well plate,
therefore regular tips are used to transfer solution from regular plate
into NMR tubes. But, it of course cannot be used to wash NMR tubes.
Occasionally, it fails to push solution to the bottom of NMR tubes.
Generally, we are very satisfied with it.

Hope this would help.

Jiangli

-----Original Message-----
From: Wheeler, Patrick
Sent: Tuesday, March 28, 2006 11:09 AM
To: Yan, Jiangli
Subject: FW: NMR liquid handler


Hi Jiangli,

Can you offer any help to these folks?

Patrick Wheeler
 
Pfizer La Jolla Labs
Open Access Team
858-526-4849 Office
858-395-2023 Mobile
 

-----Original Message-----
From: Christopher Singleton [mailto:cas40_at_bu.edu]
Sent: Monday, March 27, 2006 9:06 AM
To: ammrl_at_ammrl.org
Subject: NMR liquid handler

Greetings AMMRLers,


             I'm looking for suggestions on a liquid handler for NMR
tubes, one that will take a well plate, then dilute it up and inject it
into the NMR tube, say for eight at a time. We really don't have the
funds for a flow probe (which is the obvious solution), but we do have
several autosamplers, like a Gilson 215 or a Zinsser Lissy 2002. Has
anyone successfully fitted an autosampler for automatic filling/washing
of NMR tubes? Thanks,


 

 
Chris

----------------------------------------------------------------------
LEGAL NOTICE
Unless expressly stated otherwise, this message is confidential and may be privileged. It is intended for the addressee(s) only. Access to this E-mail by anyone else is unauthorized. If you are not an addressee, any disclosure or copying of the contents of this E-mail or any action taken (or not taken) in reliance on it is unauthorized and may be unlawful. If you are not an addressee, please inform the sender immediately.