120 Hz spike problem on our Varian INOVA 500 MHz was solved

From: Zhou, Zhe (ZZ) <"Zhou,>
Date: Wed, 23 Mar 2005 16:54:07 -0600

Hi Friends:

 

The 120 Hz spike problem on our Varian INOVA 500 MHz was solved. It was
found with scope that PTS output had 120 Hz modulation caused by a filter
capacitor. The capacitor was replaced.

 

I really appreciate the following people's suggestions and help.

 

David Redwine, Dow Chemical

Richard K. Shoemaker, University of Colorado at Boulder

Jerry Hirschinger, Purdue Interdepartmental NMR Facility

Dale Treleaven, LSU at Baton Rouge

Rolf Tschudin, National Institutes of Health

James Breeyear, UVM

Robert Harker, University of Florida

Jincheng Xiong, Dow Chemical

Lamy Chopin, Dow Chemical

Jason Burgess,RTI International

Greg Young, UNC

Josh Kurutz, U. Chicago

Corey Liu, Stanford

David Jones, University of Colorado Health Sciences Center

Perry J. Pellechia, University of South Carolina

 

Have a good day !

 

Zhe Zhou, Ph.D.

Research Specialist (NMR)

Analytical Sciences, Dow Chemical Company

2301 N. Brazosport Blvd., B-1219

Freeport, TX 77541

Tel: 979-238-1387

Fax: 979-238-0752

Email: zzhou_at_dow.com
Received on Wed Mar 23 2005 - 17:01:24 MST