--
Dr Phil Dennison
NMR Facility Director (949)824-6010 (office)
Department of Chemistry (949)824-5649 (lab)
University of California, Irvine (949)824-8571 (fax)
Irvine, CA 92697-2025 dennison_at_uci.edu
USA
~~~~
Since B-11 is a quadrupolar nucleus, the relaxation times are extremely
short. I have measured them in the low msec range. One user had a very
unusual sample that probably formed micelles and he had an amazingly long
T1 of ~50 msec as I recall. We use a relaxation delay of 2 microseconds
and a data acquisition time of 33 msec. With these parameters we always
use at least 1,000 scans, which gives a total experiment time of 44
seconds.. (Depending on the spectrometer system overhead, you may have a
minimum value for the relaxation delay that will work at all.) Because of
the short relaxation times, you of course have relatively broad lines.
(BF3-etherate is an exception due to the symmetric environment of the B11.
It?s width at half-height is ~7 Hz.) Because the lines are broad, we do
not lock, spin, or shim.
Dr. Jane Strouse
Director of Instrumentation
Department of Chemistry and Biochemistry
1410 Molecular Sciences Building
UCLA
Los Angeles, CA 90095-1569
~~~~
We have been doing some water treatment improvements to our house well
water. The goal is to remove the Borom from the current 40 PPM to less than
1 PPM so we don't kill plants we water with the well water. In this effort
we have been running NMR of water samples looking at Boron using our old
NT360 with a 12mm tube. We can easily see Boron in the 1 to 40 PPM range
running 4096 scans with a total recycle time of 0.1 sec and a 90 degree
pulse. We do not lock and observe excellent sample to sample reproducibity.
Our major problem is the background Boron signal from the Boron in the glass
NMR tube and insert. We run a blank sample of pure water and subtract it out.
Woody Conover II
woody_at_acornnmr.com
~~~~
My experiences in running 11B have been on GE 300 MHz and an Inova 500
MHz NMR systems. We found better results on our 500 MHz NMR system than
on a 300 MHz NMR system; quadrapolar effects actually decrease with field
strength. You can run a very short recycle delay with 11B; your delay
values should be fine. You may have too little boron to observe. One
major watchout; if you are using a standard 10 mm BB probe (the same as
the one I have) then you will get a strong background signal from the
probe (likely from borosilicate glass inserts in the coil area). My
solution to solve this was to increase my ringdown delay to around 1 msec.
You can verify that this is or is not a problem by collecting the
spectrum with no NMR tube present and see if any of your peaks are still
present. More elegant solutions involve using some type of echo
sequence but I never tried that. The linewidth of 11B is fairly broad;
good luck getting this to work.
Mike Krejsa
Michael R. Krejsa, Ph.D.
Manager, Analytical
Cytec Surface Specialties
730 Worcester Street
Springfield, MA 01151 USA
(413) 730-2470 Phone
(413) 730-2196 Fax
~~~~
A few comments to your Boron problem. A repetition time of 0.1 sec. is
fine. Lines are broad, but T1 is equaly short. Measuring with external
standard is not necessary, you could better use the newest varian
userlib setref, and use this as external reference. You need of course a
good pw90, but the decoupling calibration as used in 13C or 31P is good
for 11B as well. When the obeserved signals are small, the probe
background- and tube-signal must be huge. Normally your signals should
be much narrower as the background signal. We use a macro setlp (see
attachment) to minimise these signals with backwards linear prediction.
Just try setlp(n) with n=3-30 and wft. It is a bit tricky but with a
little bit experience it works fine. We measured a lot of B11 1-D and
2-D spectra with our 5mm probes on Unity-500.
good luck,
Jan runsink
"********************"
"setlp - Backward LP"
"********************"
"****************************************************"
"remove baseline_roll when alfa too large with :
"setlp(>n) lsfid=-n ; choose 'n' to minimize lp_value"
"default value: n=3 or `sw/12500` "
" if lsfid < 0 --> default=default-lsfid"
"****************************************************"
setlimit('lsfid',64000,-64000,1)
parlp
lpopt='b' lpfilt=16 lpnupts=128
while ((lpnupts+0.5) > np/2 ) do
lpfilt=lpfilt/2 lpnupts=lpnupts/2
endwhile
if ($# < 0.5) then
if (sw > 35000) then
$1=trunc(sw/10000+0.5)
else
$1=3
endif
on('lsfid'):$lsfid
if ($lsfid > 0.5) then
$lsfid=lsfid
endif
if ($lsfid < 0) then
$1=$1-$lsfid
endif
endif
if ($1 > 0.5) then
strtlp=$1 strtext=$1 lpext=$1
proc='lp'
fpmult=0.5 lvl=0 tlt=0
else
proc='ft' fpmult='n'
endif
~~~~
www.invista.com/email_disclaimer.shtml
This archive was generated by hypermail 2.4.0 : Sat Jun 10 2023 - 17:13:18 MST