Nitrogen Fill Question

From: Hoeglund, Adrienne <AHoeglund_at_aoc-resins.com>
Date: Mon, 28 Feb 2005 14:15:02 -0600

Hello. Last week I was very involved in a lively conversation concerning
proper Nitrogen fill techniques and determined that there are several
different methods for properly filling the magnet. I was wondering, for
those of you filling a 300 MHz spectrometer, with an Oxford magnet, what
fill technique do you use?
 
Here is what I was told is the absolute proper way I should be doing it. I
would like some input and/or comparisons to your methods.
1. Lower the pressure in the Nitrogen dewar to no more than 10 psi.
2. Place the two rigid transfer lines in the magnet and begin the fill
slowly.
3. Continue filling until the magnet is full.
4. Allow the system to boil-off for about 30-minutes and then top off the
system.
5. Remove the rigid tubes and put everything up.
 
>From what I was told in my conversation last week, the technique we were
using could have ruptured the Nitrogen vessel and damaged the magnet.
However, with this new technique I can't tell that liquid is being
transferred, the pressure in the Nitrogen dewar keeps dropping as the fill
is progressing and, I just stopped the fill and rather than filling the
magnet, the Nitrogen level has dropped a centimeter.
 
>From what I can tell, the assumption is made that there is a constant
pressure of gas pushing the liquid Nitrogen like there is for the helium.
And, the dewars that we get for Nitrogen, that isn't going to work.
 
Therefore, I would be interested in knowing if you do the fill as described
above or if you have some other technique. If you do use the above
technique, please fill me in on the details.
 
If there is interest, I will post a summary. If you don't want your
response posted, please let me know that as well.
 
Thank you in advance for your input.
 
Adrienne
 
Adrienne B. Hoeglund
NMR Specialist
AOC, LLC.
950 Highway 57 East
Collierville, TN 38017
(901) 854-2800 ext. 532 - work
(901) 853-1582 - fax
www.aoc-resins.com

----------------------------------------------------------------------------
------------
The information contained in this message and or attachments is intended
only for the person or entity to which it is addressed and may contain
confidential and/or privileged material. Any review, retransmission,
dissemination or other use of, or taking of any action in reliance upon,
this information by persons or entities other than the intended recipient is
prohibited. If you received this in error, please contact the sender and
delete the material from any system and destroy any copies.

----------------------------------------------------------------------------
------------

 
 
Received on Mon Feb 28 2005 - 16:47:29 MST

This archive was generated by hypermail 2.4.0 : Sat Jun 10 2023 - 17:13:18 MST