From: Hasan Celik <hcelik_at_berkeley.edu>
Date: Thu, 29 Jun 2023 19:57:27 -0700

Hello Martha,

I had a similar problem after upgrading to 4.3.0 recently. It sounds like
the routing in topshim1d2h is not properly defined between 2H amp, TRX
boards, and the amps/preamps. It is probably going through the BB amp
instead. Did you run expinstall after installing 4.1.3?

If expinstall doesn't work, I think you can also manually override this by
loading the topshim parameters via rpar topshim1d2h, manually define the
routing in edasp and save, and wpar a new parameter set. Topspin won't
allow you to overwrite the default topshim1d2h parameters so you will have
to save it under a different name and move/overwrite the folder using the
terminal. The new parameter folder should save under
/opt/topspin4.1.3/exp/stan/nmr/par/user and then you rename the folder to
topshim1d2h and move it one folder up to par. The owner for the parameter
folder should be nmrsu for the user and the group under linux so that
topspin can read that properly.

Hope this helps.

Best,

Hasan

On Thu, Jun 29, 2023 at 6:37 PM Martha Morton <mmorton4_at_unl.edu> wrote:

> Dear group,
>
> I'm running Topspin 4.1.3 on a 600 NEO console. The operating system is
> CentOS 7. Every time I update the distro topshim breaks. Initially, I saw
> errors in permissions to use shim files. Now I'm seeing errors where
> topshim will start then stop throwing the error "setupHardpulseP1plw1 - F=
or
> logic channel 1 power level not determined ".
> Has anyone seen and solved this?
>
> Thanks,
> Martha
>
> Sent from my Verizon, Samsung Galaxy smartphone
>
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#266): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/266__;!!PvDODwlR4mBZyAb0!XG28lm__OwZ-tbbdqBpCXvB2rADO8ZtG=
Y8q3h_bp_ldtZ-XJpoz3NH0Nms4mUiGqyXY5ApSN5RkREYqjFA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/99865246/=
7559972__;!!PvDODwlR4mBZyAb0!XG28lm__OwZ-tbbdqBpCXvB2rADO8ZtGY8q3h_bp_ldtZ-=
XJpoz3NH0Nms4mUiGqyXY5ApSN5Rnxrx7NWw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Jun 30 2023 - 06:04:30 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:42:24 MST