Re: [AMMRL] Pulse sequences for RDCs?

From: Eugene DeRose via groups.io <derose=niehs.nih.gov_at_groups.io>
Date: Mon, 12 Jun 2023 14:50:19 +0000

Hi Savanah,

We have used the BioPack gNhsqc_IPAP and gNhsqcS3 experiments to measure
HNRDC's of proteins. For larger proteins, the gNtrosyS3 experiment may
work better. There is also an IPAP option in the BioPack sofastNhmqc experiment.

To measure CH couplings, you might try the CTgChsqc_JCH experiment, though
we have not run this experiment.

Best regards,
Eugene

Eugene DeRose, Ph.D.
NIEHS NMR Laboratory Manager
Contractor to NIEHS/DIR
984.287.3572
derose_at_niehs.nih.gov<mailto:derose_at_niehs.nih.gov>
--


From: <main_at_ammrl.groups.io> on behalf of "Shumaker, Savanah (NIH/NCATS) [C] via groups.io" <savanah.shumaker=nih.gov_at_groups.io>
Reply-To: "main_at_ammrl.groups.io" <main_at_ammrl.groups.io>
Date: Monday, June 12, 2023 at 9:55 AM
To: "main_at_ammrl.groups.io" <main_at_ammrl.groups.io>
Subject: [AMMRL] Pulse sequences for RDCs?

Hello All,

Does anyone know of pulse sequences for measuring RDCs on Varian/Agilent
systems, using VnmrJ 4.2? Thanks in advance!

Savanah Shumaker, M.S. [C]
NMR Chemist
Analytical Chemistry Core
Division of Pre-Clinical Innovation (DPI)
NIH/NCATS
9800 Medical Center Drive
Rockville, MD 20850
Email: savanah.shumaker_at_nih.gov<mailto:savanah.shumaker_at_nih.gov>
Cell: 920-360-2642





-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#211): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/211__;!!PvDODwlR4mBZyAb0!V8rFJysgAMfy1xQcIwJmqlQLFqKRdTxi=
d_GxmxI9AuFpjxhcuSSsp_sGGgd9dWbCH3KlOJZPYEX_Q1pc98i-wkmacvMc$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/99483471/=
7559972__;!!PvDODwlR4mBZyAb0!V8rFJysgAMfy1xQcIwJmqlQLFqKRdTxid_GxmxI9AuFpjx=
hcuSSsp_sGGgd9dWbCH3KlOJZPYEX_Q1pc98i-wsLep79I$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Jun 12 2023 - 07:51:58 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:42:24 MST