i remember i change also on a pts250 (or pts160 ? ) .it was a bruker msl400
i put capacitor ouside of pts because it was too big
At 19/06/2014 15:17, Matthias Findeisen wrote:
>Dear all,
>
>I want to let you know, that all people who wrote to me, were right:
>the 100 Hz-modulation was due to the 10.000 µF electrolytic capacitors of
>the power module.
>
>Replacing them solved the problem.
>
>Thank you again.
>
>
>Yours
>Matthias Findesien