RE: AMMRL: VnmrJ 2.3A Plotting Problem

From: Thomas K Weldeghiorghis <thomaskw_at_lsu.edu>
Date: Mon, 21 Jun 2010 20:28:26 -0500

Dear AMMRL,

Thank you all for your help. We now have our HP LaserJet 6MP printer working with VnmrJ 2.3A. I uninstalled all printers using the unistall command in /vnmr/pinters/bin before I reinstalled makePrinter and created new plotter and printer following the easy instruction for makePrinter utility. The only different thing I did between now and my previous unsuccessful use of makePrinter to solve our problem is that I might not have uninstalled all printers before reinstalling makePrinter the first time around.

Thanks you all again for your help.

Regards,

Thomas

Thomas K. Weldeghiorghis, Ph.D.
NMR Facility
Department of Chemistry
Louisiana State University
232 Choppin Hall
Baton Rouge, LA 70803



-----Original Message-----
From: Thomas K Weldeghiorghis [mailto:thomaskw_at_lsu.edu]
Sent: Tue 6/15/2010 10:29 AM
To: ammrl_at_ammrl.org
Subject: AMMRL: VnmrJ 2.3A Plotting Problem
 
Dear AMMRL:

We have VnmrJ 2.3 running on RH Linux 5.4 (on a Dell Precision T3500) and we are having problem plotting. All we get when we try to plot spectra is pages and pages of text. We have HP LaserJet 1320n and HP LaserJet 6MP. I will appreciate any suggestions you might have to solve this problem.
Thanks,

Thomas

Thomas K. Weldeghiorghis, Ph.D.
NMR Facility
Department of Chemistry
Louisiana State University
232 Choppin Hall
Baton Rouge, LA 70803



-----Original Message-----
From: Hugo Azurmendi [mailto:azurmend_at_vt.edu]
Sent: Thu 6/10/2010 3:06 PM
To: ammrl_at_ammrl.org
Subject: AMMRL: are cryoprobes good on old magnets? Answer: YES
 
Dear colleagues,
thanks a lot for the prompt replies. I got about 20 answers in this
short time and no negative experience relative to the magnet, except may
be a couple that had to work extra to get a good shim, and another that
had to put new legs.
Thanks again,

Hugo Azurmendi
NMR Spectroscopist
Chemistry - Virginia Tech
Received on Mon Jun 21 2010 - 15:31:00 MST

This archive was generated by hypermail 2.4.0 : Fri Jun 16 2023 - 17:02:30 MST