Nitrogen Safety Warning

From: <stberger_at_rz.uni-leipzig.de>
Date: Sat, 8 Jun 2002 17:49:27 +0200

Nitrogen Safety Warning

Due to cost reasons, the liquid nitrogen is produced at some
universities by small liquifier machines. The quality with respect
to oxygen content of this nitrogen varies, of course, dependent
on liquifier performance and operator skill between 0.1 and 5 %.
Whereas this is probably not important for everyday laboratory
use, the consumption of this type of nitrogen in long hold
dewars, e.g. NMR magnets, may pose a problem.
Mass spectrometric analysis of the content in seven magnets at
our place, which were uninterrupted filled by this type of
nitrogen since years, indicated last week a nitrogen/oxygen ratio
approaching 1:1 (!). We are currently debating on the most
efficient and safe way to remove the problem.