Re: direct detection of nitrogen

From: Charles G. Fry <fry_at_bert.chem.wisc.edu>
Date: Wed, 28 Jun 2000 15:26:33 -0500

Wei,

Directly observing N15 at natural abundance is a very tough
proposition. I believe some students here have managed it, but
takes a LOT of sample. A 10mm (or bigger) probe would be a big
help.

I have over the last year or so done a bit of "discovering"
about doing 15N at natural abundance using inverse PFG techniques.
These experiments pull natural abundance 15N into a more normal
realm of research with regard to sample requirements. 15N with
a directly bound proton is quite easy to observe. I would
recommend normal HMQC even on a broadband probe without gradients
for these compounds before attempting direct-observe 15N (haven't
tried it, though) . The S/N improvements if it works are just to
big to not try.

Getting to 15N with long range protons is quite tricky; I
would be very surprised if these nitrogens can be detected in
any general sense (ie., on anything other than perhaps a few
samples with some special characteristics, such as extremely
high solubility) without inverse techniques. Solvent issues
(solubility and polarity?; DMSO and CD2Cl2 seem to be best) are
also crucial to the success of these experiments. And PFG is
a huge help. All my efforts in these regards follow closely on
the published work of Gary Martin and co-workers; see his
publications for many demonstrations of natural abundance 15N.

Cheers,
Charlie


At 11:41 AM 6/28/00 -0500, Wei Li 448-7532 wrote:
>Dear all,
>
>Thank you very much for the many suggestions for my question. I will try
>those suggestions with a standard sample first on our BBI probe (the only
>one we have to do X nuclei besides C13) to see if it is feasible.
>
>Regards
>
>Wei Li
>-----------------------------------------
>Wei Li
>Dept. of Pharmaceutical Sciences
>University of Tennessee
>847 Monroe, Room 327
>Memphis, TN 38163
>
>Phone 901-448-7532 (Office) Fax 901-448-6828 (Office)
>Phone 901-527-6784 (H)
>


----------------------------------------------------------
Charles G. Fry, Ph.D. Tel: (608)262-3182
Director, MR Facility Fax: (608)262-0381
Chem. Dept., Univ. Wisconsin
Madison, WI 53706 USA email: fry_at_chem.wisc.edu
----------------------------------------------------------
Received on Wed Jun 28 2000 - 17:16:26 MST

This archive was generated by hypermail 2.4.0 : Sat Jun 03 2023 - 15:53:26 MST