Hi Eugenio,
Sometime after the fill, if the pressure in the tank was too high, we got
the similar error. What we did was wait a minute or two after the refill
then cap it up. that solve the problem.
Jennifer
On Wed, Jul 19, 2023 at 10:43 AM Kenneth Sharp-Knott <kknott_at_vt.edu> wrote:
> You may have an ice blockage in your LN2 dewar ports. You can possibly
> open it up and clear the block before it goes into an irreversible warmup.
>
> Best
>
>
>
--