Dear Dr. Heffron,
Am Montag, Juli 10, 2023 12:44 CEST, schrieb "Heffron, Gregory J." <gregory_heffron_at_hms.harvard.edu>:
> Hi All,
>
> I write to ask the group if anyone knows how best to stop an acquisition
> in progress on Bruker NEO systems. We are somewhat familiar with commands to
> do this, but have not had much success in getting them to work. Specifically,
> we would like to be able to stop an acquisition in progress, then start it
> at a later time from the point it was stopped, OR be able to stop an
> acquisition, then restart it by calculating appropriate delays and add the
> data together when the total experiment is finished. We ask about this because
> we suffer occasionally from interruptions to experiments with cryoprobes
> due to cooling water issues.
>
these troubles do not belong to NEO also to the AVANCEs before!
You can 'halt' an 1D acquisition to pump up the number of scans or transients -
then you can add the freshly set numbers to the previously halted acquisition -
go is the magic word - NEVER use 'ZG' - this is zero go - then all the previously
added stuff has gone!!!
To do this with 2Ds ... no way to add some more rows or transients. You have
to start from scratch.
OK - if there is something broken in the connection - with the NEO you can
address the 'broken' stuff from the previous running acquisition - there is
one chance to get these from the EPU. You can move these data fragments on
your harddrive and then you can try your best. EPU is an enhancement to the
previous IPSO box or even the previous CCU-unit.
Good luck and have a nice try!
Yours sincerely,
Ulrich Haunz
> Any ideas are greatly appreciated, and we will post a summary.
>
> Best,
>
> Greg
>
> Gregory J. Heffron | Director, Biomolecular NMR Facility | Visiting Scientist,
> Dana-Farber Cancer Institute | Department of Cancer Biology
> Harvard Medical School 240 Longwood Ave. | Boston, MA 02115 | Fax :617-432-4383
>
>
=========================
Ulrich Haunz, Dipl.-Ing. (FH)
NMR-Spektroskopie
Raum L 710, L 807
NMR-Center M520, M521, M527
Postfach M 715
Fachbereich Chemie
Universitaet Konstanz
78457 Konstanz
Tel.: +49 (0)7531 88-4261 -> M521 Bruker Avance NEO 800
Tel.: +49 (0)7531 88-4422 -> L807 JEOL Lambda 400
Tel.: +49 (0)7531 88-4955 -> L710 Bruker AVIII HD 400
Tel.: +49 (0)7531 88-4956 -> M520a Bruker AVIII 600
Tel.: +49 (0)7531 88-4957 -> M527 Bruker AVIII 400 Automation
Tel.: +49 (0)7531 88-4958 -> M520 Bruker AVIII 400 WB
Tel.: +49 (0)170 9153213
FAX : +49 (0)7531 88-3898
=========================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#288): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/288__;!!PvDODwlR4mBZyAb0!WMbcg-_0VmT-AiuzhUaVAreBFIOuxt0k=
3rW9phjy3R0V_bhnfA9owX8yoJwIu4bgjr9pqp36A5jvcN6vXHLeamcsOxUT0g$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/100057679=
/7559972__;!!PvDODwlR4mBZyAb0!WMbcg-_0VmT-AiuzhUaVAreBFIOuxt0k3rW9phjy3R0V_=
bhnfA9owX8yoJwIu4bgjr9pqp36A5jvcN6vXHLeamcJ9ymUlg$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Mon Jul 10 2023 - 09:04:08 MST