Yep. The login field is in the service page of the bsms about halfway down
the page. Username doesn't matter but must be at least 3 characters.
The super-secret password is 1964. This will give you access to the helium
calibration buttons.
Cheers,
Mike
> On Jul 4, 2023, at 10:31 AM, Akien, Geoffrey <g.akien_at_lancaster.ac.uk> wrote:
>
> Pretty sure you need to log in as administrator in the ha pages?
>
> -----Original Message-----
> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Karel Klika
> Sent: Tuesday, July 4, 2023 4:07 PM
> To: main_at_ammrl.groups.io
> Subject: [External] [AMMRL] AMMRL: helium meter calibration
>
> Dear all,
>
> I am having trouble working out how to do the calibration of the helium meter
> for our 400s. For the 400nb, going to "ha" the helium calibration is greyed out
> while on the 400wb I cannot even access ha.
> Both systems run TopSpin 3.6 under Centos 7. (Paravision for the wide bore.)
>
> Our 600 is running TopSpin 2.1 under Windows XP and I have no problems doing the
> calibration of the helium meter.
>
> Any suggestions ?
>
> Many thanks.
>
> Kind regards,
> Karel
>
>
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#274): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/274__;!!PvDODwlR4mBZyAb0!S7DekEk-3kZX6P-MlQQrlWxLoS9pcF-U=
AwV9p1zGd_u6LF0tzElwwCJNg36hRBjPGS5E4I3OTVXwa5N4mg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/99950043/=
7559972__;!!PvDODwlR4mBZyAb0!S7DekEk-3kZX6P-MlQQrlWxLoS9pcF-UAwV9p1zGd_u6LF=
0tzElwwCJNg36hRBjPGS5E4I3OTVWEN87cVQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Jul 04 2023 - 09:39:43 MST