Re: AMMRL: compiling problems

From: George Gray WCR <george.gray_at_varianinc.com>
Date: Sat, 17 Jul 2010 08:57:45 -0700

Hi Susan,

In going to Linux the compiler requires that the "abort" statement be
replaced by "psg_abort". This problem is
encountered when people use code run on a Sun whose compiler is less
"fussy".

Regards,
  George Gray
  NMR Applications

On 7/16/2010 1:17 PM, Susan Nimmo wrote:
> Everyone:
>
> One of the groups is trying to set up a new experiment and they are
> running into some problems. I have pasted the problem below. If
> anyone has any ideas about this, we would appreciate it.
> Thanks,
> Susan Nimmo
>
>
> I am trying to install a macro to run a J-HMBC experiment. I found
> the macro and instructions for installing it on the NMR Pulse Sequence
> Library of the Carlsberg Laboratory
> (http://www.crc.dk/nmr/nmrpulse/experiments/small/jhmbc/varian/index.shtml).
> It appears that the macro was written for a system running Vnmr 6.1 on
> Solaris 7. When I run the install script, I get the following error
> messages:
>
> Compiling Sequence...
>
> Compile with DPS extensions failed
> Retrying without DPS extensions
> Compiling Sequence...
>
> Pulse Sequence did not compile
> The following errors can also be found in the file
> /home/vnmr1/vnmrsys/psglib/gj_hmbc.errors:
>
> gj_hmbc.c: In function 'pulsesequence':
> gj_hmbc.c:312: error: too many arguments to function 'abort'
> gj_hmbc.c:407: error: too many arguments to function 'abort'
>
> When I call the macro in VNMRJ, it does not create any new parameters,
> presumably because the pulse sequence was not installed. I believe
> the DPS extensions error is related to the fact that the install
> script was created for a Solaris system and I am using it on a RHEL 4
> system. Do you have any suggestions for installing this macro? Is
> there another source for a macro for the J-HMBC experiment? Any help
> you can give will be greatly appreciated.
> Sincerely,
>
> Matt Joyner
> Postdoctoral Research Fellow
> Department of Chemistry and Biochemistry
> University of Oklahoma
>
>
>
>
> Susan L. Nimmo, Ph.D.
>
> NMR Facility Manager
> Department of Chemistry and Biochemistry
> University of Oklahoma
> 620 Parrington Oval, Room 208
> Norman, Oklahoma 73019
>
> office: 405-325-0677
> cell: 405-627-7044
> lab: 405-325-2843
>
Received on Sat Jul 17 2010 - 05:58:17 MST

This archive was generated by hypermail 2.4.0 : Fri Jun 16 2023 - 18:35:03 MST