Re: AMMRL: Tube Spinner Question Summary

From: Shaw Huang <huang_at_chemistry.harvard.edu>
Date: Mon, 12 Jul 2010 15:05:42 -0400

Dear Keith and all,

Thanks for the great summary, Keith! Since one of the respondents
mentioned about Norell's using a test result done at Harvard, I have to
clarify that that test was not authorized by me or any other faculty
members here, and the use of Harvard's name in their promotion flyers
without our consent had drawn Harvard Corporate lawyers' great
attention. The student who did the test was not even aware of Norell's
intention to use his result for such purpose.

Best,
Shaw


On 7/9/2010 10:57 AM, Keith Brown wrote:
>
> Dear All:
>
> Wow! I think I struck a chord asking these questions. I can't thank
> you enough for your replies.
>
> The concensus seems to be that
>
> 1: low-field rated tubes and disposable tubes can usually be used
> satisfactorily
> in high-field magnets provided that they are not bowed and fit in
> the spinner
> turbines properly.
>
> 2. don't use a shim to keep the tube from slipping
>
> I should have mentioned in my original message that the particular
> tube
> in question was very definitely of a smaller diameter than most of the
> tubes used here and wouldn't fit properly into any of our spinners.
> 4.8 mm as opposed to an average of 4.9 mm. I don't think that the
> o-rings need
> replacing yet.
>
<editorial snip, wpn>