RE: AMMRL: Chemagnetics 1H Amplfier problems SUMMARY

From: Jerry Hirschinger <jerryh56_at_purdue.edu>
Date: Fri, 9 Jul 2010 12:01:15 -0400

Hello again,
Wow, what a great response I got! There are several issues I will detail
below in no particular order, but first let me thank Ed Twehouse, Kevin
Goehring, Woody Conover, Richard Hullihen, Marcus Giotto, Paul Kanyha, Deane
McIntyre, and Weiguo Hu for there responses and suggestions.

1. There are 2 versions of 1H amps - one in blackface made by Creative
Electronics, and one in blondface by Chemagnetics. My questions were
specifically regarding the Creative amp.
2. Creative Electronics is now Amplifier Systems, Inc. www.ampsystems.com
It is reported that they will sell a replacement power supply and amp for
less than $5,000.
3. Tomco makes a solid state amp that will replace the tube amp. Price
through Tecmag is $17,000.
4. Dust blocking the tube ventilation path can shorten tube lifetime.
5. Large electrolytic capacitors in the high voltage power supply go bad
with age. They are difficult to find on the market, but may be special
ordered with high minimum and long lead time. If your capacitors are bad,
see item 2 above.
6. Cathode bias zener diode goes bad. This was the first problem we had,
and I installed a heat sink on the new zener.
7. Cathode bias fuse F1 (1 amp) blows. See REVEAL below for more on this.
8. RF Parts <http://www.rfparts.com/tubetran.html> may be a better place to
buy the tube than Richardson Electronics.
9. New tubes should be burned in. Varian method uses 1pncyc pulse program.
Start with 50 watts of peak power, tune/match the amplifier then let it
pulse for 5 minutes at each increasing pulse width starting at 10 us, and
increasing width by a factor of 10 until 2s is reached. Repeat this
sequence at each power level in 50W steps until 500W is reached.
Alternative method (unapproved) is to leave the power supply on to heat the
tube overnight.
10. Check and properly adjust the filament voltage. Test points are on the
rear of the amp, and the adjustment is on the rear of the power supply.
11. Install a circulator/isolator on the amp output to protect the tube.
Varian Inc. sells one as part no. MISOL40000. I did not get a price. This
is a commom RF/microwave component and can be purchased from many places; be
sure to get the proper frequency and power ratings.
12. The high voltage is derived unregulated from the AC line voltage, so a
flakey AC line will give problems.
13. Always operate the amp with a load. Connect the load before unblanking
the amp or turning on the RF input. Never disconnect the probe or load
during pulsing.

Now, for the REVEAL! On our system I found 3 problems. First, the cathode
bias zener was bad. I installed a heat sink on the replacement. Second,
the cathode bias fuseholder was bad; this was a great surprise, but the
spring in the bottom of the fuseholder was stuck so the fuse had no pressure
on it. Finally, the 1H connector on our 4mm probe had come loose from the
probe body. All together it made one big pain in the butt, which I will
sooth with a few beers this weekend.
Cheers, -Jerry
Jerry Hirschinger, NMR Instrumentation Specialist
Purdue Interdepartmental NMR Facility
560 Oval Dr. West Lafayette, IN 47907-2084
Office: Wetherill 365A
Phone / Fax: (765) 494-5288 / 494-0239
Cellular: (765) 427-3034




> Hi Spinheads,
> We have a Chemagnetics Infinity 400MHz console with their 500 Watt
> high power proton amplifier. The amplifier is eating up $1,000
> triodes fairly regularly, with less and less time between failures.
> Does anyone have any maintenance suggestions to extend the life of
> these tubes? Does anyone know of a good replacement for this amp?
> Other than the tubes wearing out, are there any other maintenance
> issues you are having with the amp or its power supply? We have had
> to replace the cathode bias zener diode in the power supply, but now
> have a heat sink mounted on it to try and avoid it happening again.
Received on Fri Jul 09 2010 - 06:01:18 MST