AMMRL: pH probes that fit 5mm tubes [SUMMARY]

From: Pellechia, Perry <pellechia_at_mail.chem.sc.edu>
Date: Thu, 30 Jul 2009 08:53:15 -0400

The responses to my question was quite numerous and also very varied. I
will try to give a summary of the options and experiences presented to me
and give a rating either good/fair/poor based on the comments sent.

Almost everyone agreed that the NMR sized pH probes are very fragile but
many feel with care they are a reasonably reliable option. The general
consensus was that if you do not break them first you can expect them to
last about 18 months before they need to be replaced.

Thanks again to everyone that sent me their comments.

----------------------------------------------
At least three votes were for using a standard pH probe (6mm OD) and larger
volumes (1.5mL) with transferring into an external vial to make the pH
measurements. These were also votes against using any of the pH probes
designed for NMR tubes due to the fragility of these probes.

----------------------------------------------
3 comments on: Mettler Toledo type:6030-M3/180/1m/BNC combination pH
electrode sold by Wilmad (cat. no.. 6030-02-BNC) {GOOD}

----------------------------------------------
I was pointed to the NMRWIKi that has two Accumet Microprobes listed but no
real evaluations. One other response commented on ACCUMET probes: {GOOD}

http://nmrwiki.org/wiki/index.php?title=Micro_pH_electrodes

----------------------------------------------
One comment on: Thermo Scientific Orion "Micro Combination pH Electrode for
NMR Cells", catalog No. 9826BN {GOOD}

----------------------------------------------
One comment on: Aldrich pH electrode (Sigma-Aldrich, z113441-1ea) {GOOD}

----------------------------------------------
One comment on: Solid state probes.
Disadvantage: you need to buy a different meter.

http://www.phmeters.com/

NMR probes:

http://phmeters.com/shopsite_sc/page6.html
http://phmeters.com/shopsite_sc/page17.html
http://phmeters.com/shopsite_sc/page3.html
{UNKNOWN}

----------------------------------------------
One comment on: IQ Scientific (www.phmeters.com) with stainless steel ISFET
pH probe (PH47-SS) {GOOD <-> FAIR: expensive maybe not as accurate as a
glass probe}

----------------------------------------------
One comment on the Hamilton Spintrodes (like we are using) {GOOD}

----------------------------------------------
*****Other Options suggested*****

----------------------------------------------
Fiber optic pH sensors. These have only been around for a few years and are
still rather expensive but are flexible and generally small enough to fit
down a 5 mm NMR tube. Here are three possible sources for fiber optic pH
meters and probes:

pH-Optica from World Precision Instruments: http://www.wpiinc.com

PreSens Precision Sensing GmbH:
http://www.presens.de/products/brochures/category/sensor-probes/brochure/ph-
microsensors.html

Ocean Optics: http://www.oceanoptics.com/Products/phsensor.asp

{VERY INTERESTING}

----------------------------------------------
Brian Sykes published a paper recently on buffer components to determine pH
Internal pH indicators for biomolecular NMR.
J Biomol NMR. 2008 May;41(1):5-7

We describe a non-invasive technique for determining pH in biomolecular NMR
sample using buffer components (formate, tris, piperazine, and imidazole) as
internal pH indicators, whose (1)H NMR chemical shifts are sensitive to pH
in a range from 2.5 to 9.8. This method is suitable for a wide range of
applications where samples are handled intensively during NMR titrations or
in high throughput analysis in structural genomics or metabolomics.

----------------------------------------------



Perry.
-- 
============================================