Please see below for summary. Thank to all who responded.
Our Original Question to AMMRL
The 600MHz Varian 7700-CS Cold Probe system on our Inova 600 has been
working for about a year. Among other factors, the 7700-CS Closed-
Cycle Cryogenic System boasts a new robust CryoBay.
We would like to get feedbacks from those who own the Varian 7700-CS
Cold Probe system (or Bruker equivalent). Do you have a Service
Contract for the whole cryo system or just the Cold Head? What
problems are you experiencing?
Our understanding is Cold Head maintenance (cleaning) implies
complete Head replacement. That being the case, why not just operate
the system for 2 years or until the Head fails? It doesn’t cost any
more – we checked. And if it hasn’t failed after two years, then the
Cold Head and the Oil Absorber can be serviced at the same time. The
down side of this scenario is the Probe must be returned to Varian.
Institutions that are able to absorb being without the Cold Probe for
a month or so could potentially save a bunch of money. A full Service
Contract costs about $30,000.00 dollars/year.
** We have a correction to our Cold Probe "premise." Clarification is
important:
If you run the cold probe to the end of the lifetime of the cold head,
it will NOT require returning the probe to Varian for repair. They
will just replace the cold head on site. A failure due to lifetime of
the cold head should not damage any other part of the system according
to the vendor.
1. I think the biggest issue is “the possibility of long term
contamination”. Running the coldhead longer than recommended before
replacing it carries the risk that somehow the system may get
contaminated. It is not clear what that means exactly, but it can
happen to either a Varian or Bruker system. Removing the contamination
if it happens is tricky and also expensive. I would strongly
recommend the cryoprobe service contract. I am not a fan of console
service contracts, but I think the cryo one is money well spent. One
probe failure plus the annual coldhead service can easily exceed the
price of the contract. I think that the 7700-CS system is now
superior to the Bruker one. From what I know it is more robust, and
less likely to fail.
2. We have Bruker cryoprobes on 600, 800 and 900 systems.The annual
service costs $12,500 and includes replacing the coldhead, adsorber in
the compressor and sometimesan in-line helium filter. This service
usually takes just 3-4 daysand we never send the probe out unless
there is a serious problem.While it is tempting to take the view that
one should let these systems run beyond a year without service I think
you riskdamage to other very expensive units. Imagine if the sleevein
which the coldhead resides is damaged, lots of $$$$ to fix. We have
never had a service contract on our cryosystems.But for $30,000 it is
not to bad if that includes the annual service,parts and repair all
year and coverage of big components like
the probe, He compressor and the cryoplatform. In an ideal world we
too would probably pay for the service contract. We have our fingers
crossed that nothing serious goes wrong with our probes.We had the 600
CP rebuilt and total cost including installation and check-out in
house by an engineer was about $23,500. We had a big problem with the
800 CP one year after install. Luckily that was covered by warranty
or else I imagine that would have cost
$25K-$30K to fix.
3. We have a Bruker 600 with a cryoprobe from 2003 or thereabouts.
We don’t have any kind of maintenance contract. We run it until
something fails, then we pay for it to be fixed. It usually fails
about once a year, so then we send it in for its “annual
maintenance” (the cost of the maintenance may vary depending on what
has failed, but the PO descriptions I looked up make no such
indication). The annual maintenance costs are rising a wee bit more
than inflation:
2006: $8750
2007: $12500
2008: $12500
2009: $16575
Doubling every four years, almost. Hopefully this trend of averaging
a 19% annual increase will not continue (1.19 ** 4 = 2.01). The last
failure we experienced was with water/ice accumulating in the helium
loop. Other failures we have experienced have resulted from
compressed air interruptions (low pressure, no pressure), which messes
up the VT regulation required to keep the sample from getting too cold
(due to its close proximity to the chilled coil around the sample).
Good gas supply and good power supply are critical. We have a great a/
c so that has never been an issue.
4. I have a Bruker 500MHz TCI cryoprobe. I don't have a contract, but
pay $12,500 for the annual maintenance when it is due. It is every
10,000 hours of operation. The maintenance takes about 1.5 days and
covers cold head change, plus anything else necessary. Some units are
swapped every other service. This seems like better value than
Varian's $30K.
5. We are in the process of purchasing a cold probe and are facing the
same dilemma. I was considering the third choice of having coverage on
the cryo probe but not the chilling components. With the exception of
the probe itself, most of the components are "off the shelf" (a Google
search of "sumitomo mri compressor parts" comes back with about one
thousand hits). However, like the most of Varian's service contracts,
it is not the cheapest way to go, but it does minimize system down
time. Who can afford to be without a cryo-probe for a month or more
whenever the system feels like failing?
6. We have an older Varian Cold probe, installed in 2005. Our first
year had a lot of repairs. We don't have a service contract on it.
Pieces of the system need to be maintained on a regular maintenance
schedule-vacuum pumps, cold head, water filters,.... Maintenance isn't
hard. My suggestion is to watch Varian do the first maintenance round
and talk to the engineer about maintenance times. Then you can do much
of the maintenance yourself. Since I didn't attend ENC this year, I'm
looking for information on older systems. Mainly, how long do the
compressors last? We talked about this at ENC 3 years ago.
7. I was told that beyond 2 years you risk damaging the whole system.
Of course, if you replace the adsorber (which I think I can do myself)
you would not have that problem. I was told by the Varian engineer
that we could do the entire cold head replacement ourselves. Then I
talked to Ed Manzanares and he said the cost would be the same whether
we do it ourselves or the Varian engineer does it. So apparently it
is free for Varian to send out an engineer!
8. It is my impression that the key issue in preventing contamination
is changing the oil absorber, which I don't think requires changing
the cold head.
9. We have a Bruker Cryoprobe. Their specified maintenance is 10,000
hours or once a year. We are letting it run until it becomes a
problem, then hiring the service on a case by case basis. We never
have service contracts for anything. Our cost for Bruker’s service is
~$12,000 per 18 months. IMO, you won’t be able to predict the time of
operation between services, nor will you be able to reach 2 years on a
coldhead.
10. We have a new (fourth generation - 'PLUTO') salt tolerant 13C
enhanced cold probe that was installed in August 2008. We have the
7700 Cryobay and it is air cooled. Apart from our own internal power
interruptions it has run faultlessly and has impressive S/N (>6500:1
on a thin walled 0.1% ethylbenzene at probe temp of 20K). We are
monitoring the probe stats and I could send you them but basically the
CLI is still very low (~0.15-0.17) and only when this rises (to 3 or 4
I believe) is it time for a cold head change. We have a service
contract for 4 years that was negotiated when we did a multi-
spectrometer purchase. What is the oil absorber? I know the external
compressor system needs some annual maintenance in terms of an oil
change (or something..!) - is this what you mean?
11. One fear I would have of running the coldhead to the point where
it losses a lot of efficiency would be how much wear/extra strain
you're getting on the yoke and other mechanical linkages. We went
through several periods of ratcheting where the coldhead 'piston'
starts to stick (Bruker system - another long story). This led to
several coldheads failing - meaning they broke at a linkage below the
yoke. Bruker replaced the coldheads free of charge because they could
find no signs of contamination, but several of the failed coldheads
were really hard to remove. If a coldhead gets really stuck the
engineers warned me it could mean sending the coldbox away. At the
time I was also concerned that coldhead failure could damage the
smooth surface inside the coldbox cylinder. I'm not an expert on these
systems and ratcheting seems an extreme case, but the thought of
mechanical damage to the CCC keeps me a wake at night. Our woes with
the ratcheting turned out to be a faulty compressor by-pass valve
(still not cheap). It might be worth asking Varian how likely damage
could be and who would bear the cost of repairs.
12.Summary from the Cold Probe Discussion at AMMRL’s meeting (48th
ENC) These notes depict each major component of the cold probe/
cryoprobe systems.(Thank you Martha Morton for providing the summary)
1. Vacuum pumps: The Varian turbo pumps, that are on Generation
1 and 2 probes, should have the TIP seals changed every six months.
Otherwise they will start pulling the load and will gradually wear
out. Ion pumps are very reliable and have lifetimes of around 5 years
if used properly.
2. Cold Head (CCC): Use grade 6 helium (highest purity available)
and watch how you change tanks. The molecular sieve bed can be
contaminated causing blockages that will reduce cooling performance.
Cold head service can be pushed until the cold head dies at the risk
of several months of down time, because engineers aren’t readily
available. Plan ahead! MD10’s can go 2 years or more.
3. Cryobay Helium Compressor: It is recommended to change the
charcoal filters for oil scrubbing about every 6 months on Leibert
compressors. They have an approximate lifetime of about 5 years under
optimal working conditions. Both Varian and Bruker have moved to
Sumitomo helium compressors/cold heads on their latest cryogenic
systems.
4. Electrical Outages: Several people have bypasses until backup
generators start. Put the cold probe on emergency circuits and use
UPSs for sites that have lots of electrical storms.
5. Heat exchangers and water chillers– Copper tubing on Haskris
chillers will corrode over time specially if DI water is used. Opt
for stainless steel if possible. Many people are using water-water
heat exchangers. Filters in this case need to be monitored
carefully. Water–air heat exchangers can work well depending on your
site.
6. Boiloff from liquid N2 of the magnet can be used as your
emergency eject air. See Jerry Hirshinger’s diagram.
7. $$$$$: Most of us didn’t project the long-term costs of
running these probes. Cold head service and probe repair are
expensive and you have to plan your budget accordingly. The general
consensus was to push vendor maintenance back, but keep logs on your
system.
13. We've got two cold probes with the original style cryogenic
systems. I think the cost of the annual maintenance visit plus the
cost of a service contract on the probe only is only slightly less
than the cost of a full service contract. I don't know if you want to
run your cold head until it fails. If you monitor the heater wattage,
you can usually see it start to "droop" (give a lower value); this
means the helium gas going into the probe isn't as cold as it used to
be, so the heater doesn't have to work as hard to maintain the temp.
at 25 K. Once it drops about 1 W (or if it goes under 4 W), it might
be a good idea to think about doing the "annual" maintenance. I quote
"annual" because I once ran a cold head a little more than 20 months
(it was our 1st cold probe, and I was a little ignorant about the
maintenance issues) before the droop became significant; a service
engineer, in a moment of candor, once told me the cold heads can
probably safely go about 15 months without any worries, but after
15-18 months you should really be monitoring the performance (heater
wattage) and thinking about replacing the cold head. So the "annual"
maintenance could really be 15-18 months, EXCEPT the tip seal in the
scroll pump that provides the backing pressure for the turbo pump
usually won't last that long. The shortest interval I've had between
tip seal changes has been 6 months which is a bit of a pain. (I
haven't compared all the intervals, but my gut feeling is the average
is about 9 months, but not 12.) Varian Vacuum claims they've improved
the material in the tip seal so it'll last longer; we'll see how long
the last two tip seals I put in last.
Received on Tue Jul 07 2009 - 11:25:56 MST