AMMRL: running topspin 1.3 on a 'modern' PC update

From: Tom Pratum <tkp9551_at_gmail.com>
Date: Tue, 12 Jan 2021 14:25:26 -0800

Hello Again Everyone-

As always, I had many very excellent responses to my query. The short
answer is that, yes, TopSpin 1.3 can be run on an OS more up to date
than Windows XP. It does not appear that it will run on any other
version of Windows (with some caveats), but it will run on a variety of

Linux OSs:
- most commonly CentOS 5.x; (not the most recent CentOS release, but quite
a bit more up-to-date than Windows XP).
- not as commonly CentOS 7.x; it appears to be a bit of a project to get
that to work, and there could definitely be issues.
- a few users of RedHat 3.x and 4.x (there might be some question as to
whether these are more up-to-date than Windows XP).

Note that not all spectrometers running TopSpin 1.3 can function with
Linux installed - mostly it appears to depend on the ECL of the CCU
board (for CCU10, ECL 06 or 22).

There is an issue in switching OS’s in that the existing Bruker
TopSpin license will not work for the new version of TopSpin required,
requiring the purchase of a new software license (unless the
spectrometer service license is used).

For their very useful responses, I would like to thank:
Ta-Chung Ong, Linda Okerlund, Joe Dumais, Jeff Walton, Clemens Anklin,
Jerry Hu, José Martínez, Rainer Haessner, Josh Kurutz, Bob
Berno, Andrew Fowler, Peter Scherer, John Decatur and Jon Webb.


Tom Pratum
Southern Oregon University


> On Jan 7, 2021, at 1:10 PM, Tom Pratum <tkp9551_at_gmail.com> wrote:
>
> Hello and Happy New Year Everyone-
> We have a older Avance instrument that will not run on anything higher
> than TopSpin 1.3. It appears that we will need to replace the current
> PC, which is running Windows XP, soon. My question is: can TopSpin 1.3
> run on other operating systems? Is anyone out there running it on
> anything more recent than Windows XP (Windows 7, Windows 10, or any
> version of Linux)?
>
> Thank you in advance for any advice.
>
>
> Tom Pratum
> Southern Oregon University
Received on Tue Jan 12 2021 - 12:25:15 MST