Re: AMMRL: help with Vnmrj

From: Ron Crouch <roncrouch1_at_mac.com>
Date: Thu, 28 Jan 2016 15:18:03 -0700

Hello Camil,

All you need is a probe file in place holding all calibrations. That can
be built fully automatically using tools in a variety of places. In
Experiment selector look for Calibration. Pick what you desire. They
are grouped by calibration samples.


After that EVERYTHING you run will be properly parameterized. Even
things water suppression will be automatically setup (presat, wet,
dpfgse_water etc) and those passed with zero effort into all subsequent
experiments. 1D 2D or 3D.

Absolutely everything is easily configurable in the experiment menus.

By the way to my knowledge VnmrJ 4.2 was the last release.

Best regards,

Ron

On 1/27/2016 5:25 PM, L Joubran wrote:
> Hi all,
>
> I’ve recently adopted an old Agilent system running Vnmrj 4.4. I’ve been a Bruker user for the last >15yrs and am not used to anything else. Is there any way to get the software to behave more like Topspin? Every time I load a new experiment, I have to set all the parameters manually. Is there a way to alter the given experiments in the experiment menu? In Topspin, you load a parameter set and move ahead with gpro.
>
>