AMMRL: RE: vnmrj under Fedora: Summary

From: hsin <hsin.wang_at_sci.ccny.cuny.edu>
Date: Fri, 28 Jan 2011 18:54:11 -0500

I wish to thank Eugenio Alvarado, Andrew Fowler, and David VanderVelde for
responding to my

Question. They all pointed out that you can make links of libtcl8.3.so and
libtk8.3.so to the libtcl and libtk in current OS, in the following fashion,

 

ln -s /usr/lib/libtcl8.5.so /vnmr/lib/libtcl8.3.so
ln -s /usr/lib/libtk8.5.so /vnmr/lib/libttk8.3.so

 

And 8.4 - 8.6 are backward compatible with 8.3. This is indeed the case and
my problem solved.

 

Hsin

 

  _____

From: hsin [mailto:hsin.wang_at_sci.ccny.cuny.edu]
Sent: Wednesday, January 26, 2011 5:15 PM
To: 'AMMRL List'
Cc: ealva_at_umich.edu
Subject: AMMRL: RE: vnmrj under Fedora

 

I installed vnmrj2.2C under Fedora14 according to the procedure that Eugenio
Alvarado posted on Dec 1, 2009. VNMRJ works fine, but when I tried to
launch 'makePrinter', vnmrWish fails. The error message:

 

#/vnmr/tcl/bin/vnmrWish: error while loading shared libraries: libtk8.3.so:
cannot open shared object file: No such file or directory

 

Has anyone encountered such a problem and solved it?

BTW, has anyone installed VNMRJ under ubuntu?

 

Hsin

--
Hsin Wang, Ph. D.
NMR Facility Manager
City College of New York, CUNY
Marshak-1217
160 Convent Ave.
New York, NY 10031
212-650-5831 (phone)
212-650-8719 (fax)
 
Received on Fri Jan 28 2011 - 13:54:19 MST