AMMRL: Gradient echo spike noise

From: Zhang, Weixing <Weixing.Zhang_at_STJUDE.ORG>
Date: Wed, 20 Jan 2010 10:35:20 -0600

Hello Everybody



We have a 600 MHz HCN triple resonance probe which was used primarily for proteins in the past. Recently we used it to do small molecule screening experiments for drug discovery. Since we use 90% H2O, we have to use either water-gate or excitation sculpting for water suppression. We often see spike noises at the baseline in these experiments.



In order to find the origin of the noise, we tested many parts in the NMR console, such as transmitter, receiver, etc., but we did not find any problems. Finally, I came up with the following method to test the probe.



Sample: DSS in D2O

1H Pulse sequence: d1 --- 90---grad --- 300us ---180 ---grad ----300us---Acquisition



When the gradient level is 0, I did not see any noises in the spectrum.

When the gradient level is 1000 or higher, I saw big spike noises at the baseline of water and the methyl group of DSS.



I also tested a Nalorac probe using the same method. For the Nalorac probe, I did not see spike noises at the baseline of water and only saw a small spike at the methyl group of DSS when the gradient level is much higher.

Using the Nalorac probe, I did not see baseline spike noises in either water-gate or excitation sculpting experiments.



Has anybody seen this before?

What might be the problem with my HCN triple resonance probe?



Thanks,

Weixing Zhang