HNC-TOCSY-CH experiment on Bruker NMR

From: Robert Peterson <peterson_at_mbi.ucla.edu>
Date: Wed, 10 Jan 2001 14:16:56 -0800 (PST)

Hi everybody,

I've been having a problem getting an experiment to run on a Bruker DRX
system. It's the HNC-TOCSY-CH experiment from Art Pardi's lab (JACS vol.
118 1996 pp5316-5317). I programmed it myself so there could be something
wrong with the pulse sequence, but I think it's right.

This experiment has two hetero-TOCSYs (H to N and N to C, both DIPSI-3)
and a homonuclear TOCSY (FLOPSY-8 on carbon). It seems like there's some
instrument problem. Usually when I try to run the experiment it crashes
within 30 or 40 transients with the error message:

GCU error code 0x80:
on the fly calculation: not enough time
error during internal timing handling

Despite this message, there isn't really any problem with the gradients.

If a particular dataset once produces this error message, then it always
produces it no matter what I change.

It's possible, by reducing the number of dummy scans in a "clean" dataset
(one that hasn't yet crashed), to get it to not crash. It then will run
through, but doesn't generate any signal.

I've talked to one other person who tried this experiment on a Bruker
system and got the crash with the same error message. And the only people
I've heard of that got this experiment to work did it on Varians. It
almost seems like our systems are somehow not able to make all those TOCSY
pulses or something. I'm not trying to dis Bruker or anything. I love
Bruker! But I'm just wondering if there is a small technical difficulty
here.

Has anyone out there ever tried this experiment on a Bruker system? Any
suggestions?

Thanks!

-Robert

----------------------------------------