Re: bruker data

Charles G. Fry (fry@chem.wisc.edu)
Fri, 05 Feb 1999 14:08:28 -0600

Graham,

I have written notes that are posted on our web site for AVANCE to VNMR
conversion:

http://cic.chem.wisc.edu/nmr/Guides/BAVUG-TOC.html

The early instructions are specific to conversion of data acquired at
NMRFAM converted for our facility, but after that the notes are applicable
for anyone. You will need lsbru from the vnmr userlib. lsbru left
shifts the data and does appropriate 1st order phase correction to
getting around the Bruker digital filtering.

Is there a way in XwinNMR to get around having to do lsbru (ie. a command
that accurately removes the filter)? Someone at Bruker once told me there
was an XwinNMR command to do this, but I have been unable to find it.

I cannot state that the conversion I have posted is bullet proof for all
types of data. We've tried only a few data sets, but on those it seemed to
work ok.

Cheers,
Charlie Fry

At 08:20 AM 2/5/99 SAST-2, you wrote:
>Dear All
>
>I need to get nmr data from a Bruker DRX system in Britain to South
>Africa and process it using VNMR 5.1.
>
>Question:
>
>1. What is the best method of transferring the data? tar, compress
>and ftp?
>
>2. Our MACLIB has a conversion macro for Bruker AM and AMX systems.
>Is there a DRX macro?
>
>Thanks
>
>Graham
>
>
>Graham E Jackson
>Associate Professor
>Department of Chemistry
>University of Cape Town
>Cape Town
>South Africa
>
>Fax: (027 21) 6897499
>Tel: (027 21) 6502531
>E-mail: jackson@psipsy.uct.ac.za
>http://www.uct.ac.za/depts/cem/academic/gjackson.htm
>
>

----------------------------------------------------------
Charlie Fry Tel: (608)262-3182
Director, MR Facility Fax: (608)262-0381
Chem. Dept., Univ. Wisconsin
Madison, WI 53706 USA email: fry@chem.wisc.edu
----------------------------------------------------------