AMMRL: Topspin 1.3 and Vista - anyone?

From: Yael S. Balazs <balazs_at_technion.ac.il>
Date: Wed, 12 Feb 2014 08:12:46 +0200

Greetings to spinland,

As we continue to exorcise XP from our intranet . . . I have a DPX200
limited to Topspin 1.3 (AQX/AQR rack etc.).
My CCU09 ECL is 55, which is the minimum to run with Linux (yay -
we're lucky enough to have the appropriate hardware for this option!)
- so I'll probably go that direction.

BUT - Has anyone gotten TS 1.3 to run with Vista?
"Bruker BioSpin has not tested TopSpin 1.3 with Windows Vista and
therefore we do not support this combination."

I got as far as diskless, flexlm, and updates all installed.
The error message I get with TS1.3 is:
"A service providing data for display and processing is not available."
"The address stated below does not conform to protocol standards:
URL= (Original message = `CORBA_ILLEGAL_PROTOCOL`URL=)"
  etcetera.

I do have telnet installed on vista. All the services listed in the
instruction manual are running.
Any success stories and ideas would be most appreciated!

As a side note, the ONLY reason not to put in Linux is for consistency
across the spectrometers . . . the other two Avance I's don't have the
CCU/10 ECL compatible for Linux and so they're already running Vista.


Cheers!
Yael

-- 
Yael S. Balazs, Ph.D.                   http://nmrlab.technion.ac.il
NMR Facility Manager                    phone: +972-4-829-3748
Schulich Faculty of Chemistry           fax: +972-4-829-5703
Technion - Israel Institute of Technology
Haifa, Israel 32000                     email: balazs_at_technion.ac.il
Received on Tue Feb 11 2014 - 20:12:53 MST

This archive was generated by hypermail 2.4.0 : Sun Jun 18 2023 - 17:22:26 MST