A few questions for those of you bold VJ pioneers. My system is a
Merc Plus 400, vnmrJ 1.1D on Solaris 9, Blade 1500. After lots of ups
and downs in the course of repair, it's clear that user accounts need
to be updated. Some selections were missing from my own walkup account.
(1) After altering the system configuration as vnmr1 to select
Protune present, updating my walkup user account caused a check box
not normally there prior to be there and checked. The result will be
automatic tuning before each acquisition. I do not want this. Is
there a place in user_templates where I can at least remove the "yes"
checkmark? Or should I disable Protune as vnmr1 and only re-enable it
when something other than proton or carbon is needed? My users aren't
too swift and I see trouble ahead if they're changing probe tuning
all the time.
(2) Although I went through the procedure for calibrating Protune
seemingly successfully according to Chapter 2.7 of a Protune manual,
it doesn't work quite right. The dip moves to the target frequency,
but then instead of refining the match, it runs away from the tune
frequency a bit. I abort the process and manually move the tune and
match into position with the servo knobs. Is there a trick to getting
Protune to behave just perfectly instead of almost?
Bill