AMMRL: Summary of responses to Graduate Lab course in NMR

From: Jeff de Ropp <jsderopp_at_ucdavis.edu>
Date: Thu, 15 Feb 2007 12:38:22 -0800

Dear AMMRL,
below is my original query plus all responses on the question of samples to
use for a graduate-level NMR class for organic chemistry students.
Jeff de Ropp, UC Davis

  ------original query-------
Dear AMMRL,
For first year graduate students specializing in organic chemistry our
Chemistry department has a lecture/lab course that covers various
spectroscopic methods, with a very heavy focus on NMR. The lab is devoted
primarily to NMR, with hands-on instruction in use of the spectrometers at
300-600 MHz. The students learn to run various 1D and 2D sequences on
standard samples. We like to complement this with the students running
some unknown samples with these methods and analyzing the spectra, but we
haven't been very satisfied with how this has worked out.

So here is the query: can people send me suggestions on samples that work
well for students at this level to run and analyze by 1D and 2D NMR?
Alternatively, are there other sorts of lab exercises that students in this
group (organic chemists) and at this level could benefit from, something
more challenging than just running standard samples?

------- Responses-------

We used to use each of the 8 carbon alcohols as unknowns for training class
assignments. Also estrone methyl ether offers several issues to explore via
noesy and various 1D assignments. A nearly saturated solution of menthol
with just a few drops of dmso-d6 works very well for an introduction to
inadequate. The T1 is only 300ms so you can pulse fast. The concentration
is high so the s/n builds quickly. Analysis of the result is straightforward.
-----

A professor in my department runs an Organic Anlysis class identical to
what you describe, but it's for upper division undergrads and Master's
level grad students. On the NMR end, standard 1D type of experiments are
done, but then some more complex 2D are also done- here's the list:
1H: standard 1D
13C: standard 1D & DEPT (45, 90 & 135)
2D: COSY, HSQC, HMBC, NOESY & TOCSY
A wide range of samples are chosen and changed every year with varying
results. Samples from small, simple compounds to tetrapeptides are used and
some flexible while others rigid- all depending on the desired affect. I
don't have a complete list of all the compounds used unfortunately, but
could get one if you like.
In another (P Chem) class I run a VT 1D 1H experiment on DMF for dynamics
calculation. I run it so the students don't bake the probe and quench the
magnet.
All experiments are done on a Bruker DRX400 NMR spectrometer.
----
Obviously depends on there earlier level of the level of theory/prac 
they've had at undergrad level and how long you want them to spend on this 
exercise, but an easy one (to develop elementary shimming/acquisition/1D/2D 
skills) in moderate field magnets is 2-aminofluorene. Well dispersed 
aromatic protons at 400MHz, long range couplings to challenge shimming etc. 
Shouldn't take them very long to solve the structure though : ), I use it 
as a standard exam question for senior undergrads.
--------
You might want to introduce shift prediction ... I know many people use 
ACD, I bought NMRPredict through MestRec, and really like it.  I have heard 
of people who used the fullblown ACD structure elucidation package at that 
level and were very happy with how the students took to it.
---------
I'd be interested in a summary. I once used propylbenzene for a class - not 
very exciting. I also had students play with acquisition times and 
relaxation delays to explore the benefits of optimizing parameters.
-------
I teach NMR lab class for grad students here in the Philippines.  It's a 
small class, so they get to use hands-on the sole NMR instrument available 
here, a 400-MHz JEOL Lambda.  What I do is either of the following: assign 
per student/group an unknown sample (currently, my "unknowns" are camphor, 
menthol, coumarin, glucose and mannose--easily available solids dissolved 
in either CDCl3 or D2O) or a known sample (for this semester: cinchonidine 
and cholesterol, dissolved in CDCl3).  For the unknowns, the objective of 
the lab is to correctly identify their sample via 1D and 2D NMR techniques; 
for the knowns, it's to give the correct structure elucidation of their 
known sample.  Take note that all samples that are used are solids with 
medium molecular  weights (< 300 g/mol).  It's because (1) they're readily 
available, (2)  easy to dispense, and (3) don't degrade that much over time.
-----
I have just given the same type of course. Some of the unknowns we analyse 
by 1 and 2D NMR that have worked out to be successful are fairly simply 
pharmaceutical compounds such as chloroquine disphosphate and even simpler 
acetominophen. If you can get your hands on these the chloroquine 
disphophate especially makes a nice example.
Something more challenging than just standard samples is giving them an 
example of the kind of sample that you would run after a synthesis that may 
not be as pure as you would like.....it is excellent practice to try and 
identify impurities (lab solvents, grease etc.) and presence of starting 
materials and reagants possibly not removed after purification. You should 
be able to make such a mixture for analysis. At least it makes them try and 
combine their chemical and NMR knowledge in one exercise.
-------
Try octa-acetyl (slight misnomer) chitobiose for both 1D and 2D 
spectra.  It has 2 anomerics with widely separated chemical shifts along 
with diastereotopic protons and 2 connectivity net works that are 
especially suited for TOCSY, COSY, HSQC, HSQC-TOCSY and other hybrids.
------
one of my 1H, 13C-favourite samples for this level is a crude 
<http://dict.leo.org/ende?lp=ende&p=/gQPU.&search=cinnamic>cinnamic 
acid-ethylester , containing ~10% Z-isomer and a few % of aldehyde. I 
discuss shifts, couplings, integrals, H,H-COSY and hsqc with the spectra.
My 31P favourite is a solution of crude phosphoric acid-alkylesters in 
D2O(+ethanol) at pH~10, where you can find well separated signals of mono-, 
di- and higher esters of ortho-, meta- and tri-phosphates besides the 
signals of free phosphates.
-------
I often use a sample of santonin. It is an organic sample of intermediate 
size. Even at 400 MHz all 1H signals are resolved, but still quite 
challenging to assign. All carbons are resolved and the standard 2D 
sequences work very well. The ROESY is a bit challenging, so the stereo 
chemistry might not be easy to solve.
Sucrose, Anethole, Strychnine, Brucine are also good examples.
------
While I am not a professor, I would like to recommend a compound that gave 
me quite a challenge and really works the mind of the inexperienced solver 
(quite possibly the experienced one as well).  The component is 
trimethlyolpropane triacrylate
(<http://www.basf.com/rawmaterials/pdfs/TMPTA.pdf>http://www.basf.com/rawmaterials/pdfs/TMPTA.pdf) 
This one is great because you think you have it solved but with three arms 
being identical, it is just a great challenge.  It would allow you to add 
the MS side of analysis so your students could study the splitting patterns 
and match them up.
I guess it depends on how hard you want to be on your students, but this 
one is a gem.  I think I may have some around here and would be happy to 
look and send you a bit if you are interested.
-----
We don't have a lab component to the coourse.  I teach people how to do 
1Ds.  I wait on 2Ds until the students has appropriate compounds.  Though 
when I teach the class, I include interpretation of 2Ds.  My suggestions 
are to start the students with Webspectra (UC-SD).  For 2D, I think Notre 
Dame had some good compounds.  I used easy examples from research data, 
where overlap in 1H or 13C made assignment difficult.  Letitia Yao has a 
pretty nice PowerPoint presentation developed, that I might use, if I have 
to teach it again (too many instruments- not enough time).
-----------
This may be too simple, but, consider:
  0.5g menthol in d6-benzene, and,
  0.5M sucrose in D2O
-----------
I took a similar course as an undergrad, although 3/4 ths of the class was 
first year grad students.  We culminated with an unknown determination.  We 
were given several pieces of data, which included hetcor, MS and IR.  Each 
group of 4 students received the same data for their unknown.  I believe 
mine was menthanol.  We had access to some data, but not everything, which 
made the correct determination more difficult.
------
Here are some good and relatively cheap compounds readily available through 
Aldrich.  As a general rule, I try to get unknowns in the C10-15 range, 
preferably with both aromatic/olephinic and aliphatic regions.
Longifolene
L-Cinchodine
Epicatechin
Eburnamonine
Myrtanol, cis and trans
Nootkatone
3-Carene
2-Hydroxy-3-pinanone
My favorite "simple" compound is ethyl nipecotate.  It has beautiful 
dispersion.
------------
We have similar course for senior undergraduate students and first year 
posgraduate students. Also we try to challenge the students with some 
"unkown" samples and normally I try sucrose in DMSO for instanse then they 
can play around with some experiment as NOE (or NOESY) some times what we 
do, mainly when we have good students, is to ask the research labs for 
"real samples" but in this cases some times the challenges is to hard.
------------
We have been running a NMR workshop for over 5 years now primarily graduate 
students, post-docs and up.  This is the first year they are dragging us 
kicking and screaming to offer it for credit.  We do separate sessions for 
big molecule people and small molecule people.  Anyhow, we haven't used 
unknowns, but we use a Pro-Leu-Gly-amide peptide (Sigma) in d-DMSO for both 
groups.  This gives very nice spectra on 500 and 600 where we hold our lab. 
The participants run 1-D H and C, 1-D NOESY and 1-D TOCSY, 2D COSY, NOESY, 
TOCSY, HMQC, HMBC, and multiplicity edited HSQC (to compare HMQC w/ HMQC 
and see editing info.). They also do 2D 15N HSQC at natural abundance and 
learn gradient shimming. They also get a good taste of the sensitivity 
difference in the indirect detect vs. 1-D carbon.  The older peptide 
samples have picked up enough water that we do 1-D presat experiments for 
suppression and sometimes do 1-D WET experiments to suppress both the DMSO 
and water peaks in the sample.  We also make up a sample of cholesterol 
with a couple of solvents added to contaminate it to do a 1D WET 
suppression.  We usually ask people to bring a real sample from their lab 
if they have it so they see what a real sample might look like and the 
difficulties associated with getting clean data.
-----------
I have used linalool and santonin as samples for 2D NMR (200 MHz)   work 
for students to examine. I ask them to assign every carbon and   every 
hydrogen (make a table) in the molecules (I do not use these as   unknowns, 
though).
-------------
We like menthol and camphor.  Pinene gives nice spectra, but the coupling 
can be confusing (a missing 3-bond, but a strong W-coupling). A 
per-acetated glucose molecule also gave nice spectra.  We used menthol for 
a Mosher ester stereochem. determination.  I used a penicillin derivative 
for an NOE problem.
Mostly we look for samples that are floating around in people's research
groups, though sometimes it's hard to find things that are pure.
On a related issue, we have difficulty finding good sources of spectral 
problems.  The ones on the web have answers available and I don't care for 
how Crews or Silverstein present theirs (Hz/mm, what's that?), plus Crews 
basically gives the answers too.  It would be nice to have a database of 
fids, with answers only available to instructors, so we can display the 
spectra as we prefer.
--------------
We used to teach an NMR interpretation class, both in-house and for ACS. 
There are many good test molecules out there, but for a graduate course I'd 
suggest you consider:
Retrorsine
cyclopentafuranone (Aldrich 341576)
bromomethyl pyran (Aldrich 262579)
biotin
taxol actually works OK for the "final exam" kind of sample.  It's busy, 
but almost first order and they can do it.
Steroids are the perfect choice for HSQCTOXY examples.
 From the industry perspective, it's critical that they learn to do 
"structure" identification, not "NMR" or "MS" identification.  As they 
learn, give them *all* the data (or let them collect it) and let them sort 
it out.  Giving them just one type of data to interpret leads to a grossly 
skewed view of how the real job is done.
------
I'd suggest just letting students bring and make their own sample. We don't 
do this for a course, but training is more fun when they bring their own 
rather than run the same old standards. You probably couldn't run the whole 
course that way, just the last two or three weeks.
-----------
We use a number of natural product samples available from Aldrich, 18-30 
carbons, and students do a full 2D analysis (COSY, HSQC, HMBC, and ROESY or 
TOCSY) as well as 1D (1H, 13C, DEPT, 1D-NOE and 1D-TOCSY).  For 
undergraduates, we use a series of steroid unknowns and they are told that 
it's a steroid.
--------------------
I usually use ethyl nipecotate and dibucaine to showcase 2-D spectra.  The 
one problem with dibucaine is that there are 3 methyl groups, 2 equivalent 
and one that just happens to overlap, but when you notice that one triplet 
integrates out to nine protons people realize that the peaks are 
overlapping.  So far I've been pretty happy with these examples.
-------------
I use verbenol. I don't do unknowns, because it is too time consuming to 
administer. Instead, have the students do many types of experiments on this 
molecule to compare the kinds of information one can get.
Jeffrey S. de Ropp, Ph.D.
NMR Facility
UC Davis
One Shields Avenue
Davis   CA   95616
jsderopp_at_ucdavis.edu
530-752-7677 office
530-752-6480 MS1-D Lab
530-752-5682 600 MHz Lab
530-754-8238 Chem NMR Lab
530-752-3516 Fax
NMR Facility Web site URL is:
http://www.nmr.ucdavis.edu/