Dear Colleagues,
This is a follow-up to my request for pointers on 25Mg NMR. 25Mg turns
out to be a pretty good nucleus. Clemens Anklin, Alan Kenwright,
Christopher Barr, Manfred Holz and Roy Hoffman. If i missed someone,
please accept my apologies. Here are the replies recovered from my
bulging inbox:
Mg has a limited shift range of ca 20 - 30 ppm. The less symmetric the
species the broader the line. A ca 10M solution of MgCl2 is used as a
reference. The pulse width should be somewhat longer than the one you
get for 15N since the frequency is lower.
I don't have any direct experience here, but my usual standby when
faced with such a request is my battered old copy of Brevard and
Granger. It suggests sensitivity should be better and linewidths
should be narrower than for O17 (at least for symmetric species like
MgCl2 - reprted linewidth of 7Hz). They also reference the work of S.
Forsen and B. Lindman in Vol 11 of "Annual reports of NMR Spectroscopy"
(1980 or 1981?)
I tried it once on a 300 and found the signal quite easily using a Mg
salt.
I no longer have access to the data so I can't supply details. If I
remember correctly it was >200 ppm wide but still easily visible. I
have not done 17O so I can't compare, but I would put it roughly on a
par with 39K.
I made together with Prof. O. Lutz, relatively early , some work on
25Mg+ in aqueous solution (see: M. Holz, S. Günther, O. Lutz, A. Nolle,
and P.G. Schrade, Z. Naturforschung, 34a, 944-949 (1979) The
relaxation rate of the ionic nucleus 25 Mg+ in water at 25°C is 1/T1
= 3.85 s-1, that means a not too broad line compared with typical 17O
lines ! In other papers of O.Lutz, in the 80's , there might be
more 25Mg work.
25Mg-NMR is much more sensitive that 17O-NMR, with sensitivity
comparable to 13C-NMR. It also yields sharper lines in symmetric
environments as compared to 17O-NMR. For further details see
http://chem.ch.huji.ac.il/nmr/techniques/1d/row3/mg.html