Hi Rainer,
I have 5 magnets now with a total of 6 helium transfer lines. 5 of them
are nonflexible types. Once I purchased the flexible transfer line from
Bruker 2 years ago for our new system, I use it all the time to fill all 5
magnets. It seems to transfer He quite fast compared to all others I have.
I always keep He pressure below 1 psi during transfer. I also find it easy
to maneuver.
Sandip
>Hi all,
>
>we are here in a discussion to exchange our flexible helium transfer
>line for our DMX 750 by a nonflexible one. The specialists of an
>institute dealing with low temperature physics told us, that our flexible
>tranfer line wastes about 30% of helium due to the comparative high pressure
>needed to work with this flexible line.
>What's your opinion? Does anyone has experience with nonflexible transfer
>lines? Do you think 30% is a correct value?
>
>Best regards
>
>Rainer
>-----------------------------------------------------------------------------