AMMRL: Responses to and follow-ups on labile proton cross peak question

From: <weiguoh_at_polysci.umass.edu>
Date: Thu, 24 Dec 2009 10:09:20 -0500

1. My original question:

The cross peaks from labile protons in small molecules are often missing despite
strong couplings. This is seen in COSY most often. But I have not found a
convincing explanation. Proton exchange or short T2 don't seem to fully account
for it. Anybody knows of a decisive answer? Also, should we expect the same
kind of absence in NOESY?

2. Responses. There were few responses to my question. The following is a more
detailed one:

> The multiplet pattern for proton-proton coupling may not be observable
> when the lifetime for protons is less than 1/J. Omitting relaxation effects,
> the COSY cross peaks intensity is given by
>
> sin(w1t1) sin(pJt1) sin(w2t2) sin(pJt2)
>
>
> where w1 and w2 are the chemical shifts of the protons, respectively, and J
> is the coupling between the protons. t1 and t2 are the evolution time in
> the 1st and 2nd dimension, respectively. Provided t1 is not too short, it
> can be shorter than 1/J and yet the above term can still give rise to an
> observable COSY cross peak. The above equation also explains why at faster
> exchange or shorter lifetime the cross peak signal is weaker, and very weak
> peaks may not be observable.
>
> I suspect the situation would be worse in NOESY because the
> cross peak signal is much weaker than those of COSYs. If T1 is much shorter
> than the mixing time, no cross-peak signals survive and thus cannot be
> observe. Also in faster exchange the NOSY (EXSY) cross-peaks are larger.


3. I did some research myself and now I believe I can better understand it. The
absence of J-splitting and COSY cross peak for hydroxyl protons is due to
exchange of hydroxyl protons among the solute molecules. J-splitting and COSY
cross peak rely on distinct up and down population for participating protons, so
when the population goes on fast exchange, the up and down populations cancel
each other and thus splittings are not observed. The shape of the hydroxyl peak
depends on the relative amplitude of splitting strength and exchange rate.

If we infinitely dilute the sample, we should slow down the exchange and thus
expect to see J-splittings and cross peaks.

This phenomenon is most remarkable if solvent molecules are inert. If solvent
molecules are polar, they will extend labile proton life time and slow down the
exchange rate among solute hydroxyl protons.

4. I found one paper which discussed this:

Conformational Analysis. IX. Conformational and Substituent Dependence of the
Hydroxyl Proton Magnetic Resonance of Ary1carbinols’9z
Robert J. Ouellette, Douglas L. Marks, and Daniel Miller. JACS 89:4, 913

5. NOESY. If the above argument holds, then NOESY cross peak should not be
affected by proton exchange since it relies on relaxation. During relaxation,
all nuclei work toward the same direction, so the effect is not canceled even if
there is exchange among them.

Have a great holiday!

Weiguo